Methods to reduce lipemic interference in clinical chemistry tests: a systematic review and recommendations.
Soh, Sheila X; Loh, Tze Ping; Sethi, Sunil K; Ong, Lizhen.
Afiliação
  • Soh SX; Department of Laboratory Medicine, National University Hospital, Singapore, Singapore.
  • Loh TP; Department of Laboratory Medicine, National University Hospital, Singapore, Singapore.
  • Sethi SK; Department of Laboratory Medicine, National University Hospital, Singapore, Singapore.
  • Ong L; Department of Laboratory Medicine, National University Hospital, Singapore, Singapore.
Clin Chem Lab Med ; 60(2): 152-161, 2022 01 27.
Article em En | MEDLINE | ID: mdl-34773729
ABSTRACT

OBJECTIVES:

Lipemia is the presence of abnormally high lipoprotein concentrations in serum or plasma samples that can interfere with laboratory testing. There is little guidance available from manufacturers or professional bodies on processing lipemic samples to produce clinically acceptable results. This systematic review summarizes existing literature on the effectiveness of lipid removal techniques in reducing interference in clinical chemistry tests.

METHODS:

A PubMed search using terms relating to lipid removal from human samples for clinical chemistry tests produced 1,558 studies published between January 2010 and July 2021. 15 articles met the criteria for further analyses.

RESULTS:

A total of 66 analytes were investigated amongst the 15 studies, which showed highly heterogenous study designs. High-speed centrifugation was consistently effective for 13 analytes albumin, alkaline phosphatase (ALP), alanine aminotransferase (ALT), aspartate aminotransferase (AST), total bilirubin, creatine kinase (CK), creatinine (Jaffe method), gamma-glutamyl transferase (GGT), glucose (hexokinase-based method), lactate dehydrogenase (LDH), phosphate, potassium, and urea. Lipid-clearing agents were uniformly effective for seven analytes ALT, AST, total bilirubin, CK, creatinine (Jaffe method), lipase, and urea. Mixed results were reported for the remaining analytes.

CONCLUSIONS:

For some analytes, high-speed centrifugation and/or lipid-clearing agents can be used in place of ultracentrifugation. Harmonized protocols and acceptability criteria are required to allow pooled data analysis and interpretation of different lipemic interference studies.