Your browser doesn't support javascript.
loading
Gaussian-Mixture Modelling of A-Mode Radiofrequency Scans for the Measurement of Arterial Wall Thickness.
Annu Int Conf IEEE Eng Med Biol Soc ; 2021: 5598-5601, 2021 11.
Article em En | MEDLINE | ID: mdl-34892393
Measurement of arterial wall thickness is an integral component of vascular properties and health assessment. State-of-the-art automated or semi-automated techniques are majorly applicable to B-mode images and are not available for entry-level in-expensive devices. Considering this, we have earlier developed and validated an image-free (A-mode) ultrasound device, ARTSENS® for the evaluation of vascular properties. In this work, we present a novel gaussian-mixture modeling-based method to measure arterial wall thickness from A-mode frames, which is readily deployable to the existing technology. The method's performance was assessed based on systematic simulations and controlled phantom experiments. Simulations revealed that the method could be confidently applied to A-mode frames with above-moderate SNR (>15 dB). When applied to A-mode frames acquired from the flow-phantom setup (SNR > 25 dB), the mean error was limited to (2 ± 1%), and RMSE was 19 µm, on comparison with B-mode measurements. The measured and reference wall thickness strongly agreed with each other (r = 0.88, insignificant mean bias = 7 µm, p = 0.16). The proposed method was capable of performing real-time measurements.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artérias Carótidas Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artérias Carótidas Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2021 Tipo de documento: Article