Automation improves repeatability of retinal oximetry measurements.
Karlsson, Robert Arnar; Olafsdottir, Olof Birna; Helgadottir, Vedis; Belhadj, Soumaya; Eliasdottir, Thorunn Scheving; Stefansson, Einar; Hardarson, Sveinn Hakon.
Afiliação
  • Karlsson RA; Faculty of Medicine, University of Iceland, Reykjavik, Iceland.
  • Olafsdottir OB; Faculty of Electrical and Computer Engineering, University of Iceland, Reykjavik, Iceland.
  • Helgadottir V; Faculty of Medicine, University of Iceland, Reykjavik, Iceland.
  • Belhadj S; The National University Hospital of Iceland, Reykjavik, Iceland.
  • Eliasdottir TS; Faculty of Medicine, University of Iceland, Reykjavik, Iceland.
  • Stefansson E; Eberhard Karls Universität, Tübingen, Germany.
  • Hardarson SH; The National University Hospital of Iceland, Reykjavik, Iceland.
PLoS One ; 16(12): e0260120, 2021.
Article em En | MEDLINE | ID: mdl-34914738
ABSTRACT

PURPOSE:

Retinal oximetry is a technique based on spectrophotometry where images are analyzed with software capable of calculating vessel oxygen saturation and vessel diameter. In this study, the effect of automation of measurements of retinal vessel oxygen saturation and vessel diameter is explored.

METHODS:

Until now, operators have had to choose each vessel segment to be measured explicitly. A new, automatic version of the software automatically selects the vessels once the operator defines a measurement area. Five operators analyzed image pairs from the right eye of 23 healthy subjects with semiautomated retinal oximetry analysis software, Oxymap Analyzer (v2.5.1), and an automated version (v3.0). Inter- and intra-operator variability was investigated using the intraclass correlation coefficient (ICC) between oxygen saturation measurements of vessel segments in the same area of the retina.

RESULTS:

For semiautomated saturation measurements, the inter-rater ICC was 0.80 for arterioles and venules. For automated saturation measurements, the inter-rater ICC was 0.97 for arterioles and 0.96 for venules. For semiautomated diameter measurements, the inter-rater ICC was 0.71 for arterioles and venules. For automated diameter measurements the inter-rater ICC was 0.97 for arterioles and 0.95 for venules. The inter-rater ICCs were different (p < 0.01) between the semiautomated and automated version in all instances.

CONCLUSION:

Automated measurements of retinal oximetry values are more repeatable compared to measurements where vessels are selected manually.