Your browser doesn't support javascript.
loading
Wire Cutting Method Using Rotational Atherectomy for Stretched Spring Wire During Coronary Intervention.
Tanaka, Kota; Okamura, Atsunori; Iwamoto, Mutsumi; Watanabe, Satoshi; Nagai, Hiroyuki; Sumiyoshi, Akinori; Suzuki, Satoshi; Tanaka, Hirokazu; Koyama, Yasushi; Iwakura, Katsuomi; Fujii, Kenshi.
Afiliação
  • Tanaka K; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Okamura A; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Iwamoto M; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Watanabe S; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Nagai H; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Sumiyoshi A; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Suzuki S; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Tanaka H; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Koyama Y; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Iwakura K; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
  • Fujii K; Division of Cardiology, Sakurabashi Watanabe Hospital, Osaka, Japan.
JACC Case Rep ; 3(17): 1842-1848, 2021 Dec 01.
Article em En | MEDLINE | ID: mdl-34917965
ABSTRACT
During removal of an entrapped guidewire, the core wire can fracture, allowing stretching of the spring wire while the distal guidewire is still entrapped. We resolved this issue with rotational atherectomy, allowing cutting of the spring wire at the intended site, regardless of the proximal spring wire fracture. (Level of Difficulty Advanced.).