Explant Analysis of Utah Electrode Arrays Implanted in Human Cortex for Brain-Computer-Interfaces.
Woeppel, Kevin; Hughes, Christopher; Herrera, Angelica J; Eles, James R; Tyler-Kabara, Elizabeth C; Gaunt, Robert A; Collinger, Jennifer L; Cui, Xinyan Tracy.
Afiliação
  • Woeppel K; Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, United States.
  • Hughes C; Center for the Neural Basis of Cognition, Pittsburgh, PA, United States.
  • Herrera AJ; Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, United States.
  • Eles JR; Center for the Neural Basis of Cognition, Pittsburgh, PA, United States.
  • Tyler-Kabara EC; Rehab Neural Engineering Labs, Pittsburgh, PA, United States.
  • Gaunt RA; Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, United States.
  • Collinger JL; Center for the Neural Basis of Cognition, Pittsburgh, PA, United States.
  • Cui XT; Rehab Neural Engineering Labs, Pittsburgh, PA, United States.
Front Bioeng Biotechnol ; 9: 759711, 2021.
Article em En | MEDLINE | ID: mdl-34950640
ABSTRACT
Brain-computer interfaces are being developed to restore movement for people living with paralysis due to injury or disease. Although the therapeutic potential is great, long-term stability of the interface is critical for widespread clinical implementation. While many factors can affect recording and stimulation performance including electrode material stability and host tissue reaction, these factors have not been investigated in human implants. In this clinical study, we sought to characterize the material integrity and biological tissue encapsulation via explant analysis in an effort to identify factors that influence electrophysiological performance. We examined a total of six Utah arrays explanted from two human participants involved in intracortical BCI studies. Two platinum (Pt) arrays were implanted for 980 days in one participant (P1) and two Pt and two iridium oxide (IrOx) arrays were implanted for 182 days in the second participant (P2). We observed that the recording quality followed a similar trend in all six arrays with an initial increase in peak-to-peak voltage during the first 30-40 days and gradual decline thereafter in P1. Using optical and two-photon microscopy we observed a higher degree of tissue encapsulation on both arrays implanted for longer durations in participant P1. We then used scanning electron microscopy and energy dispersive X-ray spectroscopy to assess material degradation. All measures of material degradation for the Pt arrays were found to be more prominent in the participant with a longer implantation time. Two IrOx arrays were subjected to brief survey stimulations, and one of these arrays showed loss of iridium from most of the stimulated sites. Recording performance appeared to be unaffected by this loss of iridium, suggesting that the adhesion of IrOx coating may have been compromised by the stimulation, but the metal layer did not detach until or after array removal. In summary, both tissue encapsulation and material degradation were more pronounced in the arrays that were implanted for a longer duration. Additionally, these arrays also had lower signal amplitude and impedance. New biomaterial strategies that minimize fibrotic encapsulation and enhance material stability should be developed to achieve high quality recording and stimulation for longer implantation periods.