Your browser doesn't support javascript.
loading
Alterations in magnetic resonance imaging characteristics of bioabsorbable magnesium screws over time in humans: a retrospective single center study.
Sonnow, Lena; Ziegler, Andreas; Pöhler, Gesa H; Kirschner, Martin H; Richter, Maximilian; Cetin, Mustafa; Unal, Melih; Kose, Ozkan.
Afiliação
  • Sonnow L; Department of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Ziegler A; Medizincampus Davos, Davos, Switzerland.
  • Pöhler GH; School of Mathematics, Statistics and Computer Science, University of KwaZulu-Natal, Pietermaritzburg, South Africa.
  • Kirschner MH; Department of Cardiology, University Heart & Vascular Center Hamburg, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Richter M; Department of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Cetin M; Syntellix AG, Hannover, Germany.
  • Unal M; Private Practice Center Leinhausen, Hannover, Germany.
  • Kose O; Department of Radiology, Antalya Education and Research Hospital, Antalya, Turkey.
Innov Surg Sci ; 6(3): 105-113, 2022 Jan 17.
Article em En | MEDLINE | ID: mdl-35224178
ABSTRACT

OBJECTIVES:

This study aimed to examine the alterations in magnetic resonance imaging (MRI) characteristics of bioabsorbable magnesium (Mg) screws over time in a single center study in humans.

METHODS:

Seventeen patients who underwent medial malleolar (MM) fracture or osteotomy fixation using bioabsorbable Mg screws and had at least one postoperative MRI were included in this retrospective study. Six of them had more than one MRI in the postoperative period and were subject of the artifact reduction measurements. 1.5T or 3T MRI scans were acquired in different periods in each patient. The size and extent of the artifact were assessed independently by two experienced radiologists both quantitatively (distance measurement) and qualitatively (Likert scale).

RESULTS:

In the quantitative measurements of the six follow-up patients the screw's signal loss artifact extent significantly decreased over the time, regardless of the MRI field strength (p<0.001). The mean artifact reduction was 0.06 mm (95% confidence interval [CI] 0.05-0.07) for proton density weighted [PDw] and 0.04 mm (95% CI 0.03-0.05) for T1 weighted (T1w) sequences per week. The qualitative assessments similarly showed significant artifact reduction in all MRI sequences. Different imaging findings, like bone marrow edema (BME), liquid collections, and gas formation were reported. The overall inter-reader agreement was high (κ=0.88, p<0.001).

CONCLUSIONS:

The time-dependent artifact reduction of Mg screws in postoperative controls might indicate the expected self-degradation of the Mg implants. In addition, different MRI findings were reported, which are characteristic of Mg implants. Further MRI studies are required to get a better understanding of Mg imaging properties.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies / Qualitative_research Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies / Qualitative_research Idioma: En Ano de publicação: 2022 Tipo de documento: Article