Performance evaluation of novel fluorescent-based lateral flow immunoassay (LFIA) for rapid detection and quantification of total anti-SARS-CoV-2 S-RBD binding antibodies in infected individuals.
Shurrab, Farah M; Younes, Nadin; Al-Sadeq, Duaa W; Liu, Na; Qotba, Hamda; Abu-Raddad, Laith J; Nasrallah, Gheyath K.
Afiliação
  • Shurrab FM; Biomedical Research Center, Qatar University, 2713, Doha, Qatar.
  • Younes N; Biomedical Research Center, Qatar University, 2713, Doha, Qatar; Department of Biomedical Science, College of Health Sciences, Member of QU Health, Qatar University, 2713, Doha, Qatar.
  • Al-Sadeq DW; Biomedical Research Center, Qatar University, 2713, Doha, Qatar; College of Medicine, Member of QU Health, Qatar University, 2713, Doha, Qatar.
  • Liu N; Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China.
  • Qotba H; Department of Clinical Research, Primary Health Care Centers, 26555, Doha, Qatar.
  • Abu-Raddad LJ; Infectious Disease Epidemiology Group, Weill Cornell Medicine-Qatar, Cornell University, Qatar Foundation-Education City, 24144, Doha, Qatar; World Health Organization Collaborating Centre for Disease Epidemiology Analytics on HIV/AIDS, Sexually Transmitted Infections, and Viral Hepatitis, Weill Cor
  • Nasrallah GK; Biomedical Research Center, Qatar University, 2713, Doha, Qatar; Department of Biomedical Science, College of Health Sciences, Member of QU Health, Qatar University, 2713, Doha, Qatar. Electronic address: gheyath.nasrallah@qu.edu.qa.
Int J Infect Dis ; 118: 132-137, 2022 May.
Article em En | MEDLINE | ID: mdl-35231609
BACKGROUND: A vast majority of the commercially available lateral flow immunoassay (LFIA) is used to detect SARS-CoV-2 antibodies qualitatively. Recently, a novel fluorescence-based lateral flow immunoassay (LFIA) test was developed for quantitative measurement of the total binding antibody units (BAUs) (BAU/mL) against SARS-CoV-2 spike protein receptor-binding domain (S-RBD). AIM: This study aimed to evaluate the performance of the fluorescence LFIA FinecareTM 2019-nCoV S-RBD test along with its reader (Model No.: FS-113). METHODS: Plasma from 150 reverse trancriptase-PCR (RT-PCR)-confirmed positive individuals and 100 prepandemic samples were tested by FincareTM to access sensitivity and specificity. For qualitative and quantitative validation of the FinCareTM measurements, BAU/mL results of FinCareTM were compared with results of 2 reference assays: the surrogate virus-neutralizing test (sVNT, GenScript Biotech, USA) and the VIDAS®3 automated assay (BioMérieux, France). RESULTS: FinecareTM showed 92% sensitivity and 100% specificity compared with PCR. Cohen's Kappa statistic denoted moderate and excellent agreement with sVNT and VIDAS®3, with values being 0.557 (95% CI: 0.32-0.78) and 0.731 (95% CI: 0.51-0.95), respectively. A strong correlation was observed between FinecareTM/sVNT (r = 0.7, p < 0.0001) and FinecareTM/VIDAS®3 (r = 0.8, p < 0.0001). CONCLUSION: FinecareTM is a reliable assay and can be used as a surrogate to assess binding and neutralizing antibody response after infection or vaccination, particularly in none or small laboratory settings.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: SARS-CoV-2 / COVID-19 Tipo de estudo: Diagnostic_studies / Prognostic_studies / Qualitative_research Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: SARS-CoV-2 / COVID-19 Tipo de estudo: Diagnostic_studies / Prognostic_studies / Qualitative_research Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article