Comparison between the Transcutaneous and Total Serum Bilirubin Measurement in Malay Neonates with Neonatal Jaundice.
Mohamed, Mazrah; Ibrahim, Nor Rosidah; Ramli, Noraida; Abdul Majid, Noorizan; Yacob, Najib Majdi; Nasir, Ariffin.
Afiliação
  • Mohamed M; Department of Paediatrics, School of Medical Sciences, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
  • Ibrahim NR; Department of Paediatrics, School of Medical Sciences, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
  • Ramli N; Hospital Universiti Sains Malaysia, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
  • Abdul Majid N; Department of Paediatrics, School of Medical Sciences, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
  • Yacob NM; Hospital Universiti Sains Malaysia, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
  • Nasir A; Department of Paediatrics, School of Medical Sciences, Universiti Sains Malaysia, Kubang Kerian, Kelantan, Malaysia.
Malays J Med Sci ; 29(1): 43-54, 2022 Feb.
Article em En | MEDLINE | ID: mdl-35283687
Background: This study aims to investigate the reliability of the Dräger Jaundice Meter JM-105 for the screening of neonatal jaundice in Malay neonates. Methods: A cross-sectional study was conducted in a university hospital involving 130 jaundiced neonates requiring serum bilirubin determination from day 2 to day 7 of life. Results: The mean total serum bilirubin (TSB) was 232 µmol/L, whereas the mean transcutaneous bilirubin (TcB) measured at the forehead and sternum were 222 µmol/L and 223 µmol/L, respectively. Further, TcB underestimates TSB with a mean difference of 10.10 µmol/L at the forehead and 9.27 µmol/L at the sternum. A positive linear relationship was observed between TSB with TcB forehead (r = 0.82) and TcB sternum (r = 0.80). A good discriminations ability was observed for both the TcB forehead (receiver operating characteristics [ROC] curve = 89.8%) and sternum (ROC curve = 89.7%) at a TSB level of 205 µmol/L. The sensitivity ranges from 84.4% to 85.3%, while the specificity ranges from 77.4% to 76.4%. Conclusion: Our study demonstrates a strong linear relationship and good diagnostic accuracy of TcB values compared to TSB values. To conclude, TcB measured at the forehead or sternum is a good alternative as a non-invasive screening tool for non-severe hyperbilirubinemia in Malay neonates.