Early assessment of acute ischemic stroke in rabbits based on multi-parameter near-field coupling sensing.
Li, Gen; Yin, Shengtong; Jian, Man; Chen, Jingbo; Zeng, Lingxi; Bai, Zelin; Zhuang, Wei; Xu, Bingxin; He, Shengjie; Sun, Jian; Chen, Yujie.
Afiliação
  • Li G; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Yin S; Department of Neurosurgery, Southwest Hospital, Army Medical University, Chongqing, China.
  • Jian M; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Chen J; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Zeng L; College of Biomedical Engineering, Army Medical University, Chongqing, China.
  • Bai Z; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Zhuang W; College of Biomedical Engineering, Army Medical University, Chongqing, China.
  • Xu B; College of Biomedical Engineering, Army Medical University, Chongqing, China.
  • He S; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Sun J; School of Pharmacy and Bioengineering, Chongqing University of Technology, Chongqing, China.
  • Chen Y; College of Biomedical Engineering, Army Medical University, Chongqing, China. sunjian@tmmu.edu.cn.
Biomed Eng Online ; 21(1): 20, 2022 Mar 27.
Article em En | MEDLINE | ID: mdl-35346206
ABSTRACT

BACKGROUND:

Early diagnosis and continuous monitoring are the key to emergency treatment and intensive care of patients with acute ischemic stroke (AIS). Nevertheless, there has not been a fully accepted method targeting continuous assessment of AIS in clinical.

METHODS:

Near-field coupling (NFC) sensing can obtain the conductivity related to the volume of intracranial components with advantages of non-invasiveness, strong penetrability and real-time monitoring. In this work, we built a multi-parameter monitoring system that is able to measure changes of phase and amplitude in the process of electromagnetic wave (EW) reflection and transmission. For investigating its feasibility in AIS detection, 16 rabbits were chosen to establish AIS models by bilateral common carotid artery ligation and then were enrolled for monitoring experiments.

RESULTS:

During the 6 h after AIS, the reflection amplitude (RA) shows a decline trend with a range of 0.69 dB and reflection phase (RP) has an increased variation of 6.48° . Meanwhile, transmission amplitude (TA) and transmission phase (TP) decrease 2.14 dB and 24.29° , respectively. The statistical analysis illustrates that before ligation, 3 h after ligation and 6 h after ligation can be effectively distinguished by the four parameters individually. When all those parameters are regarded as recognition features in back propagation (BP) network, the classification accuracy of the three different periods reaches almost 100%.

CONCLUSION:

These results prove the feasibility of multi-parameter NFC sensing to assess AIS, which is promised to become an outstanding point-of-care testing method in the future.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Acidente Vascular Cerebral / AVC Isquêmico Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Animals / Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article