Your browser doesn't support javascript.
loading
Case Report: MR-Guided Adaptive Radiotherapy, Some Room to Maneuver.
Li, Winnie; Winter, Jeff; Padayachee, Jerusha; Dang, Jennifer; Kong, Vickie; Chung, Peter.
Afiliação
  • Li W; Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, ON, Canada.
  • Winter J; Department of Radiation Oncology, University of Toronto, Toronto, ON, Canada.
  • Padayachee J; Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, ON, Canada.
  • Dang J; Department of Radiation Oncology, University of Toronto, Toronto, ON, Canada.
  • Kong V; Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, ON, Canada.
  • Chung P; Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, ON, Canada.
Front Oncol ; 12: 877452, 2022.
Article em En | MEDLINE | ID: mdl-35494044
ABSTRACT

Background:

A magnetic resonance linear accelerator (MR-Linac) provides superior soft tissue contrast to evaluate inter- and intra-fraction motion and facilitate online adaptive radiation therapy (ART). We present here an unusual case of locally advanced castrate-resistant prostate cancer treated with high-dose palliative ultra-hypofractionated radiation therapy on the MR-Linac with significant inter-fraction tumor regression. Case Presentation The patient was a 65-year-old man diagnosed with metastatic prostate cancer to bone and pelvic lymph nodes 7 years prior. At diagnosis, he presented with a PSA of 23 ng/ml and was commenced on a luteinizing hormone-releasing hormone agonist, achieving a PSA nadir of 4.68 ng/ml at 12 months. The patient subsequently had progressive lower urinary tract symptoms, his PSA increased to 47 ng/ml, and there was a markedly enlarged pelvic mass involving the prostate with gross extra-capsular disease and invasion into the posterior bladder wall. The patient was referred for palliative radiation to the pelvic mass due to urinary symptoms, pain, and lower limb paraesthesia. Treatment was planned to be delivered on the MR-Linac with a schedule of 36 Gy over 6 weekly factions allowing for maximal target dose delivery while minimizing surrounding organs at risk (OARs) radiation exposure. Unexpectedly, the target volume had a marked 49% (453 cc to 233 cc) reduction that was accounted for in the online adaptive process. A new reference plan was generated after 3 fractions to add sacral plexus as an OAR, previously not visible due to mass encroachment. The patient reported ongoing reduction in urinary symptoms, pelvic pain, and lower limb paresthesia by the end of treatment.

Conclusion:

Using daily MR-guided ART, improved visualization of the changing target and OARs ensured safe dose escalation. The unexpected positive response of the target and improved patient outcomes demonstrated the added value of the MR-Linac for online adaptive radiotherapy in this setting.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article