Comparison between C-MAC and King Vision video laryngoscope (channelled blade) for tracheal intubation in aerosol-prevention intubation box for COVID-19 patients: A manikin-based study.
Indian J Anaesth
; 66(3): 187-192, 2022 Mar.
Article
em En
| MEDLINE
| ID: mdl-35497695
ABSTRACT
Background and Aims:
The risk of contracting infection while intubating a coronavirus disease 2019 (COVID-19)-positive patient can be reduced by the use of personal protective equipment (PPE), video laryngoscope (VL) and aerosol-preventing intubation box. We compared two VLs (C-MAC and King Vision laryngoscope [KVL]) for ease of intubation and time taken to intubate the manikin using an intubation box.Methods:
This randomised study involved healthcare workers having experience in using both C-MAC and KVL. After explaining the study and five practice sessions, a total of 63 volunteers were included; 61 participants gave consent and were enroled. The participants were allowed to intubate initially with one VL as per random sequence. Each participant performed three tracheal intubations with each device (C-MAC VL and KVL) on a manikin using an aerosol-prevention box over the head end at the time of intubation.Results:
Time taken, percentage of glottic opening (POGO) score and the number of attempts taken for successful intubation with C-MAC and KVL were comparable in any of the three attempts (P > 0.05). The participants reported more difficulty in using KVL compared to C-MAC, and insertion of laryngoscope blade into the mouth of manikin for intubation was easy in group C-MAC compared to KVL in all three intubations (P < 0.01).Conclusion:
C-MAC and KVL take comparable time for successful intubation under COVID-19 simulation conditions. But C-MAC is more user-friendly.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Clinical_trials
Idioma:
En
Ano de publicação:
2022
Tipo de documento:
Article