Determination of the Isotopic Composition of Ytterbium by MC-ICP-MS Using an Optimized Regression Model.
Anal Chem
; 94(20): 7200-7209, 2022 05 24.
Article
em En
| MEDLINE
| ID: mdl-35549212
In this study, we measure the absolute isotope ratios of ytterbium (Yb) by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) using an optimized regression model for mass bias correction. A rhenium (Re) reference material (NIST SRM 3143), which has been characterized previously, is selected as a primary calibrator to calibrate the absolute Yb isotope ratios for three Yb materials (GSB, Alfa Yb, and GBW). The three-isotope plot for all collected data indicates that the results of Yb isotope ratios obtained are not affected by any polyatomic interferences and the mass-independent isotopic fractionation. Furthermore, the recalibrated Hf historical isotope ratios by using the absolute Yb isotopic composition obtained in this study for the isobaric interference correction on Hf isotopes are in agreement with the original historical values. This work has further demonstrated the applicability of the regression model for the calibrated measurements of absolute isotope ratios using MC-ICP-MS. The three mono-elemental Yb standard solutions are thus proposed as the reference materials for Yb isotope ratio measurements in environmental and geoscience applications.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Itérbio
/
Isótopos
Tipo de estudo:
Prognostic_studies
Idioma:
En
Ano de publicação:
2022
Tipo de documento:
Article