Analysis of image quality and radiation dose in routine adult brain helical and wide-volume computed tomography procedures.
Garba, I; Fatima, A M; Abba, M; Yakubu, M; Mansur, Y; Lawal, Y; Abubakar, Auwal; Usman, Aminu U.
Afiliação
  • Garba I; Department of Medical Radiography, College of Health Sciences, Bayero University Kano, Nigeria. Electronic address: igarba.radg@buk.edu.ng.
  • Fatima AM; Department of Medical Radiography, College of Health Sciences, Bayero University Kano, Nigeria.
  • Abba M; Department of Medical Radiography, College of Health Sciences, Bayero University Kano, Nigeria.
  • Yakubu M; Department of Radiology, College of Health Sciences, Bayero University Kano, Nigeria.
  • Mansur Y; Department of Radiology, College of Health Sciences, Bayero University Kano, Nigeria.
  • Lawal Y; Department of Radiology, College of Health Sciences, Bayero University Kano, Nigeria.
  • Abubakar A; Department of Radiography, College of Health Sciences, University of Maiduguri, Maiduguri, Nigeria.
  • Usman AU; Department of Radiology, Gombe State University, Gombe, Nigeria.
J Med Imaging Radiat Sci ; 53(3): 429-436, 2022 09.
Article em En | MEDLINE | ID: mdl-35717379
ABSTRACT

OBJECTIVES:

To evaluate and compare image quality and radiation dose between the helical and wide-volume scans to determine the protocol that provides a lower radiation dose without loss in image quality.

METHODS:

The study was prospectively conducted on consented adult patients that presented for routine brain CT. Image quality and radiation dose were compared between the helical and wide-volume scans on the Toshiba 160-slice Aquilion Prime CT scanner. The volume computed tomography dose index (CTDIvol) and dose length product (DLP) for each scan mode were collected and compared. Image quality was quantitatively and qualitatively evaluated using the unenhanced brain CT images. The data were analysed using a statistical package for social sciences (SPSS) software version 20 for both the descriptive and inferential statistics. A significant difference in image quality and radiation dose between the helical and wide-volume scans was determined based on a p-value of <0.05.

RESULTS:

A total of 54 participants were included, with two groups of 27 participants. The CTDIvol and DLP values were significantly p < 0.05 higher in the helical scan (CTDIvol 65 mGy; DLP 1597 mGy.cm) compared to the wide-volume scan (CTDIvol 54 mGy; DLP 1133 mGy.cm). The grey and white matters show a better signal-to-noise ratio (SNR) for the helical scan. Meanwhile, the contrast-to-noise ratio (CNR) was significantly p < 0.05 higher in the wide-volume scan. The results from the visual grading methods were compared and showed superior image quality in helical over the wide-volume scan.

CONCLUSION:

Wide-volume provides a lower dose compared to helical and therefore, could be adopted as the routine protocol for brain CT for in house dose optimisation. Where clinical conditions warrant the need for a helical scan, the protocol should be optimised in line with the as low as reasonable achievable (ALARA) principle.