Your browser doesn't support javascript.
loading
Near Elimination of In Vitro Predicted Extrathoracic Aerosol Deposition in Children Using a Spray-Dried Antibiotic Formulation and Pediatric Air-Jet DPI.
Farkas, Dale; Thomas, Morgan L; Hassan, Amr; Bonasera, Serena; Hindle, Michael; Longest, Worth.
Afiliação
  • Farkas D; Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, Virginia, 23284-3015 , USA.
  • Thomas ML; Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, Virginia, 23284-3015 , USA.
  • Hassan A; Department of Pharmaceutics, Virginia Commonwealth University, Richmond, Virginia, USA.
  • Bonasera S; Department of Pharmaceutics, Virginia Commonwealth University, Richmond, Virginia, USA.
  • Hindle M; Department of Pharmaceutics, Virginia Commonwealth University, Richmond, Virginia, USA.
  • Longest W; Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, Virginia, 23284-3015 , USA. pwlongest@vcu.edu.
Pharm Res ; 40(5): 1193-1207, 2023 May.
Article em En | MEDLINE | ID: mdl-35761163
ABSTRACT

PURPOSE:

This study evaluated the in vitro aerosol performance of a dry powder antibiotic product that combined a highly dispersible tobramycin powder with a previously optimized pediatric air-jet dry powder inhaler (DPI) across a subject age range of 2-10 years.

METHODS:

An excipient enhanced growth (EEG) formulation of the antibiotic tobramycin (Tobi) was prepared using a small particle spray drying technique that included mannitol as the hygroscopic excipient and trileucine as the dispersion enhancer. The Tobi-EEG formulation was aerosolized using a positive-pressure pediatric air-jet DPI that included a 3D rod array. Realistic in vitro experiments were conducted in representative airway models consistent with children in the age ranges of 2-3, 5-6 and 9-10 years using oral or nose-to-lung administration, non-humidified or humidified airway conditions, and constant or age-specific air volumes.

RESULTS:

Across all conditions tested, mouth-throat depositional loss was < 1% and nose-throat depositional loss was < 3% of loaded dose. Lung delivery efficiency was in the range of 77.3-85.1% of loaded dose with minor variations based on subject age (~ 8% absolute difference), oral or nasal administration (< 2%), and delivered air volume (< 2%). Humidified airway conditions had an insignificant impact on extrathoracic depositional loss and significantly increased aerosol size at the exit of a representative lung chamber.

CONCLUSIONS:

In conclusion, the inhaled antibiotic product nearly eliminated extrathoracic depositional loss, demonstrated high efficiency nose-to-lung antibiotic aerosol delivery in pediatric airway models for the first time, and provided ~ 80% lung delivery efficiency with little variability across subject age and administered air volume.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Inaladores de Pó Seco / Antibacterianos Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Child / Child, preschool / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Inaladores de Pó Seco / Antibacterianos Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Child / Child, preschool / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article