Your browser doesn't support javascript.
loading
Less expensive, reliable blood glucose self-monitoring.
Diabetes Care ; 10(3): 357-9, 1987.
Article em En | MEDLINE | ID: mdl-3595401
ABSTRACT
We examined the reliability of split test strips for blood glucose self-monitoring. One hundred visual readings were performed with each of the following test strips original Haemoglukotest 20-800 (HGT 1/1; Boehringer-Mannheim, Mannheim, FRG) and Haemoglukotest 20-800 halved either by a splitting device (HGT 1/2 SP) or by a pair of scissors (HGT 1/2 SC). Within the plasma glucose range of 20-360 mg/dl, there was a close correlation between the results of all three test-strip readings and the values obtained by the reference method (Beckman glucose analyzer) HGT 1/1, r = .950; HGT 1/2 SP, r = .966; and HGT 1/2 SC, r = .966. Absolute deviations from the reference values were 17 +/- 2, 17 +/- 1, and 17 +/- 1 mg/dl (mean +/- SE). Analysis of variance did not reveal any significant differences between the values of each of the three procedures compared with the reference method (P = .98). These results indicate that visual reading of split test strips is as reliable as the reading of original test strips. A special splitting device is not necessary. Diabetic patients should be advised to use split test strips for blood glucose self-monitoring to reduce costs.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fitas Reagentes / Autocuidado / Glicemia / Monitorização Fisiológica Limite: Humans Idioma: En Ano de publicação: 1987 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fitas Reagentes / Autocuidado / Glicemia / Monitorização Fisiológica Limite: Humans Idioma: En Ano de publicação: 1987 Tipo de documento: Article