Your browser doesn't support javascript.
loading
Mechanical Timer-Actuated Fluidic Dispensing System: Applications to an Automated Multistep Lateral Flow Immunoassay with High Sensitivity.
Seder, Islam; Ham, Kyeong-Min; Jun, Bong-Hyun; Kim, Sung-Jin.
Afiliação
  • Seder I; Department of Mechanical Engineering, Konkuk University, Seoul 05029, Republic of Korea.
  • Ham KM; Department of Bioscience and Biotechnology, Konkuk University, Seoul 05029, Republic of Korea.
  • Jun BH; Department of Bioscience and Biotechnology, Konkuk University, Seoul 05029, Republic of Korea.
  • Kim SJ; Department of Mechanical Engineering, Konkuk University, Seoul 05029, Republic of Korea.
Anal Chem ; 94(37): 12884-12889, 2022 09 20.
Article em En | MEDLINE | ID: mdl-36069050
ABSTRACT
In this study, we present a fluidic dispensing system that can automate the sequential fluidic delivery of multiple reagents for lateral flow assays. Highly sensitive assays typically require multiple solution-based sequences, including washing steps and signal amplification. However, implementation of these types of sequences on an automated and highly sensitive point-of-care testing (POCT) platform remains challenging. Our platform consists of two disposable cartridges with reagent chambers and a test strip and an instrument that has a mechanical timer to actuate the cam-follower-gear components. The timer rotation sequentially shifts the position of the chambers and loads the reagents to the test paper strip. The dispensing intervals are controlled at a variation of <1% within a total actuation time of 60 min. Unlike other POCT devices, the timing of fluid delivery in our timer-actuated platform is not dependent on the selection of substrates and reagents, and the unique approach to fluidic delivery results in no reagent overlap or carryover, minimal reagent loss, and highly accurate fluidic timing control for highly sensitive solution-based assays. As a model application, the proposed platform applies a gold enhancement solution to amplify the detection signal and detect prostate-specific antigen with a limit of detection of 86 pg/mL within 27 min. This platform provides an opportunity for solution-based POCT applications with high sensitivity, thereby satisfying the requirement for user-friendly operations in resource-limited settings.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes Imunológicos / Antígeno Prostático Específico Tipo de estudo: Diagnostic_studies Limite: Humans / Male Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes Imunológicos / Antígeno Prostático Específico Tipo de estudo: Diagnostic_studies Limite: Humans / Male Idioma: En Ano de publicação: 2022 Tipo de documento: Article