Incidence, causes, correlates, and outcome of bioprosthetic valve dysfunction and failure following transcatheter aortic valve implantation.
Nitsche, Christian; Koschutnik, Matthias; Donà, Carolina; Mutschlechner, David; Halavina, Kseniya; Spinka, Georg; Dannenberg, Varius; Mascherbauer, Katharina; Sinnhuber, Leah; Kammerlander, Andreas; Winter, Max-Paul; Bartko, Philipp; Goliasch, Georg; Pibarot, Philippe; Hengstenberg, Christian; Mascherbauer, Julia.
Afiliação
  • Nitsche C; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Koschutnik M; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Donà C; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Mutschlechner D; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Halavina K; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Spinka G; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Dannenberg V; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Mascherbauer K; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Sinnhuber L; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Kammerlander A; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Winter MP; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Bartko P; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Goliasch G; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
  • Pibarot P; Herzzentrum Währing, Vienna, Austria.
  • Hengstenberg C; Institut Universitaire de Cardiologie et de Pneumologie de Québec/Québec Heart and Lung Institute, Department of Medicine, Laval University, Québec, Canada.
  • Mascherbauer J; Department of Internal Medicine II, Medical University of Vienna, Vienna, Austria.
Eur Heart J Cardiovasc Imaging ; 24(6): 796-806, 2023 05 31.
Article em En | MEDLINE | ID: mdl-36099163
ABSTRACT

AIMS:

Bioprosthetic valve dysfunction (BVD) is a major concern regarding transcatheter aortic valve implantation (TAVI) durability. We aimed to assess incidence, correlates, causes, and outcome of early to mid-term BVD after TAVI in relation to patient's life expectancy. METHODS AND

RESULTS:

Consecutive TAVI recipients (2007-20) with a follow-up ≥1 year were prospectively included. BVD and bioprosthetic valve failure (BVF) were assessed according to Valve-Academic-Research-Consortium-3. BVD/BVF and all-cause death served as endpoints. Average life expectancy was calculated from National Open Health Data and patients were stratified according to tertiles (1st <6.85 years, 2nd 6.85-9.7 years, 3rd >9.7 years). Of 1047 patients (81.6 ± 6.8 years old, EuroSCORE II 4.5 ± 2.5), ≥2 follow ups were available from 622 (serial echo cohort). After a median echo follow up of 12.2 months, incidence rates of BVD/BVF were 8.4% (95% confidence interval 6.7-10.3), and 3.5% (2.5-4.9) per valve-year, respectively, without differences between life expectancy tertiles. The incidence of BVD was two-fold higher within the first year of implant (9.9% per valve-year) vs. beyond (4.8% per valve-year). Valve-in-valve procedure and residual stenosis, but not age/life expectancy predisposed for BVD. BVD/BVF were independently associated with outcome for patients in the first [adjusted hazard ratio (AHR) 1.72 (1.06-2.88)/2.97 (1.72-6.22)] and second [AHR 1.96 (1.02-3.73)/2.31 (1.00-5.30)], but not the third tertile of life expectancy (P = n.s.).

CONCLUSIONS:

In this large prospective observational cohort, early to mid-term BVD after TAVI occurred at the same rate across the spectrum of life expectancy and was associated with increased mortality in patients with short but not in those with the longest life expectancy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Etiology_studies / Incidence_studies / Risk_factors_studies Limite: Aged / Aged80 / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Etiology_studies / Incidence_studies / Risk_factors_studies Limite: Aged / Aged80 / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article