Development and validation of a reversed-phase high-performance liquid chromatography-ultraviolet method for abemaciclib-related substance detection in bulk drugs.
Ni, Ronghua; Du, Xiao; Huang, Rongrong; Wu, Wenyi; Xu, Jingqiu; Ma, Xiaofei; Cai, Liangliang.
Afiliação
  • Ni R; Department of Pharmacy, Affiliated Hospital of Nantong University, Nantong, P. R. China.
  • Du X; Department of Pharmacy, Nanjing Medical Center for Clinical Pharmacy, Nanjing Drum Tower Hospital, The Affiliated Hospital of Nanjing University Medical School, Nanjing, P. R. China.
  • Huang R; Department of Pharmacy, Affiliated Hospital of Nantong University, Nantong, P. R. China.
  • Wu W; Department of quality inspection, Sinopharm Holding Nantong Ltd, Nantong, P. R. China.
  • Xu J; Department of Pharmacy, Affiliated Hospital of Nantong University, Nantong, P. R. China.
  • Ma X; Department of Pharmacy, Affiliated Hospital of Nantong University, Nantong, P. R. China.
  • Cai L; Department of Pharmacy, Affiliated Hospital of Nantong University, Nantong, P. R. China.
J Sep Sci ; 45(22): 4070-4078, 2022 Nov.
Article em En | MEDLINE | ID: mdl-36106903
Abemaciclib is an effective selective cyclin-dependent kinases 4 and 6 inhibitors for cancer therapy. The abemaciclib-related substances influence its efficacy and safety, and are important in process preparation studies and quality control. Thus, a reversed-phase high-performance liquid chromatography method was developed and validated for the detection of related substances in its bulk drug. The separation of abemaciclib and related substances was performed on a Phenomenon Gemini C18 column (4.6 × 250 mm, 5 µm) with a flow rate of 1.0 ml/min. The ultraviolet detection wavelength was 280 nm. Mobile phase A was composed of a mixed solution of aqueous solution and acetonitrile (9:1, v/v). The aqueous solution (pH 2.5) contained 0.025-mM potassium dihydrogen phosphate solution and 0.4% triethylamine. Mobile phase B was composed of acetonitrile. This novel method exhibits good system suitability, specificity, precision, stability, linearity (0.1-20 µg/ml), repeatability, and durability. Among abemaciclib and related substances, the lowest limit of detection and quantitation were 0.02 and 0.06 µg/ml, respectively, for abemaciclib. The recovery rates for related substances were above 95%. In addition, a novel degradation product was found during the process. In summary, a reliable reversed-phase high-performance liquid chromatography method was developed for abemaciclib-related substance detection in bulk drugs.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cromatografia de Fase Reversa Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cromatografia de Fase Reversa Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2022 Tipo de documento: Article