Mechanical stability of angulated zirconia abutments supporting maxillary anterior single crowns on narrow-diameter implants.
Ma, Ruiyang; Yu, Ping; Zhang, Yuqiang; Xie, Chenyang; Tan, Xin; Sun, Jikui; Yu, Haiyang.
Afiliação
  • Ma R; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China.
  • Yu P; Department of Stomatology, Chengdu Second People's Hospital, Chengdu, China.
  • Zhang Y; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China.
  • Xie C; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China.
  • Tan X; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China.
  • Sun J; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China.
  • Yu H; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd section, Chengdu, 610041, Sichuan, China. YHYmechanics@163.com.
Clin Oral Investig ; 27(1): 221-233, 2023 Jan.
Article em En | MEDLINE | ID: mdl-36161530
OBJECTIVES: To investigate the fracture strength of angulated hybrid abutments supporting anterior single crowns on narrow-diameter implants (NDIs). MATERIAL AND METHODS: Zirconia abutment with angulations of labial inclination 0° (TZ0Z), 15° (TZ15Z), 30° (TZ30Z) and palatal inclination 15° (TZ - 15Z) was designed on 3.3-mm titanium-zirconium (Ti-Zr) NDIs. Titanium abutment connected with Ti-Zr implant (TZ0T) and 0° zirconia abutment connected with pure titanium (Ti) implant (T0Z) were control groups. Thirty-six un-restored abutments and 36 abutments restored with highly translucent zirconia (HTZ) crowns were tested. Failure loads were compared among 6 groups, and bending moments were calculated for comparison between un-restored and restored abutments. RESULTS: Failure loads of un-restored abutments were affected by the abutment angle. Sixty-seven percent samples in TZ30Z and 83% samples in TZ - 15Z group fractured at the thinnest part of the zirconia abutment and exhibited lower failure load (p < .05). Failure loads of restored abutments were close to or exceeded the maximum bite force of anterior teeth, and no differences were found among six groups (p > .05). Except TZ15Z and TZ0T group, the bending moment increased with the crown construction, especially for TZ30Z and TZ - 15Z groups (p < .001). CONCLUSIONS: The fracture strength of hybrid abutments restored with HTZ crown on Ti-Zr NDIs exceeded the bite forces of anterior teeth for all the groups and were not affected by the abutment angle. CLINICAL RELEVANCE: In terms of fracture strength, Ti-Zr NDIs combined with angulated hybrid abutments and HTZ crowns can be used in the anterior region.