Inter-transducer variability of ultrasound image quality in obese adults: Qualitative and quantitative comparisons.
Maar, Megan; Lee, Juhyun; Tardi, Anthony; Zheng, Yuan-Yi; Wong, Candance; Gao, Jing.
Afiliação
  • Maar M; Rocky Vista University, Ivins, UT, USA.
  • Lee J; Loyola University Medical Center, Maywood, IL, USA.
  • Tardi A; Siemens Healthineers, Issaquah, WA, USA.
  • Zheng YY; Shanghai Jiaotong University, Shanghai, China.
  • Wong C; Rocky Vista University, Ivins, UT, USA.
  • Gao J; Rocky Vista University, Ivins, UT, USA. Electronic address: jgao@rvu.edu.
Clin Imaging ; 92: 63-71, 2022 Dec.
Article em En | MEDLINE | ID: mdl-36208612
ABSTRACT

PURPOSE:

Acquiring high-quality ultrasound images of deep abdominal organs and vasculatures in obese adults (BMI >30 kg/cm2) is considered challenging. The aim of the study was to assess the inter-transducer variability in B-mode and color Doppler image quality from four commercial ultrasound transducers through qualitative and quantitative analyses.

METHODS:

Four curvilinear transducers on three ultrasound scanners were used to acquire B-mode and color Doppler images of deep abdominal structures in 15 obesity ≥ class II (BMI >35 kg/cm2) adults. Using visual-qualitative assessment and an offline image processing software, visual-qualitative score and quantitative mean pixel values of B-mode images, and color area ratios of color Doppler images were calculated. Differences in these values among the transducers were analyzed using one-way ANOVA. The intra- and inter-observer reliability of visual-qualitative assessment and offline image processing was tested using the intraclass correlation coefficient (ICC).

RESULTS:

Differences in visual-qualitative score, mean pixel value of B-mode images, and color area ratio of color Doppler images among the four transducers were significant (p < 0.001). Transducer -4 produced the highest quality of B-mode (45-53% improvement) and color Doppler (22-73% improvement) images among the transducers. Intra-observer repeatability and inter-observer reproducibility were higher with performing offline image processing than visual-qualitative assessment (ICC 0.97-0.99 versus ICC 0.76-0.97).

CONCLUSION:

There was significant image quality variability between different transducers. Transducer -4, a transducer designed specifically for high BMI patients, had the highest quality B-mode and color Doppler images compared to the other transducers lending to improved ultrasonographic visualization in obese patients.