Your browser doesn't support javascript.
loading
Development and Validation of an Algorithm for the Digitization of ECG Paper Images.
Randazzo, Vincenzo; Puleo, Edoardo; Paviglianiti, Annunziata; Vallan, Alberto; Pasero, Eros.
Afiliação
  • Randazzo V; Department of Electronics and Telecommunications, Politecnico di Torino, 10129 Turin, Italy.
  • Puleo E; Dipartimento di Fisica, Università di Torino & Sezione INFN di Torino, 10125 Turin, Italy.
  • Paviglianiti A; Department of Electronics and Telecommunications, Politecnico di Torino, 10129 Turin, Italy.
  • Vallan A; Department of Electronics and Telecommunications, Politecnico di Torino, 10129 Turin, Italy.
  • Pasero E; Department of Electronics and Telecommunications, Politecnico di Torino, 10129 Turin, Italy.
Sensors (Basel) ; 22(19)2022 Sep 21.
Article em En | MEDLINE | ID: mdl-36236237
The electrocardiogram (ECG) signal describes the heart's electrical activity, allowing it to detect several health conditions, including cardiac system abnormalities and dysfunctions. Nowadays, most patient medical records are still paper-based, especially those made in past decades. The importance of collecting digitized ECGs is twofold: firstly, all medical applications can be easily implemented with an engineering approach if the ECGs are treated as signals; secondly, paper ECGs can deteriorate over time, therefore a correct evaluation of the patient's clinical evolution is not always guaranteed. The goal of this paper is the realization of an automatic conversion algorithm from paper-based ECGs (images) to digital ECG signals. The algorithm involves a digitization process tested on an image set of 16 subjects, also with pathologies. The quantitative analysis of the digitization method is carried out by evaluating the repeatability and reproducibility of the algorithm. The digitization accuracy is evaluated both on the entire signal and on six ECG time parameters (R-R peak distance, QRS complex duration, QT interval, PQ interval, P-wave duration, and heart rate). Results demonstrate the algorithm efficiency has an average Pearson correlation coefficient of 0.94 and measurement errors of the ECG time parameters are always less than 1 mm. Due to the promising experimental results, the algorithm could be embedded into a graphical interface, becoming a measurement and collection tool for cardiologists.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Sinais Assistido por Computador / Eletrocardiografia Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Sinais Assistido por Computador / Eletrocardiografia Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article