Near-Infrared Spectroscopy Usefulness in Validation of Hyperventilation Test.
Sandru, Stefan; Buzescu, Dan; Zahiu, Carmen Denise Mihaela; Spataru, Ana; Panaitescu, Anca Maria; Isac, Sebastian; Balan, Cosmin Ion; Zagrean, Ana-Maria; Pavel, Bogdan.
Afiliação
  • Sandru S; Department of Functional Sciences, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
  • Buzescu D; Department of Functional Sciences, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
  • Zahiu CDM; Department of Functional Sciences, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
  • Spataru A; Department of Critical Care, King's College Hospital Denmark Hill, London SE5 9RS, UK.
  • Panaitescu AM; Department of Obstetrics and Gynecology, Filantropia Clinical Hospital Bucharest, 011171 Bucharest, Romania.
  • Isac S; Department of Obstetrics and Gynecology, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
  • Balan CI; Department of Functional Sciences, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
  • Zagrean AM; Department I of Cardiovascular Anesthesiology and Intensive Care, "Prof. C. C Iliescu" Emergency Institute for Cardiovascular Diseases, 050474 Bucharest, Romania.
  • Pavel B; Department of Functional Sciences, Carol Davila University of Medicine and Pharmacy, 050474 Bucharest, Romania.
Medicina (Kaunas) ; 58(10)2022 Oct 06.
Article em En | MEDLINE | ID: mdl-36295560
ABSTRACT

Background:

The hyperventilation test is used in clinical practice for diagnosis and therapeutic purposes; however, in the absence of a standardized protocol, the procedure varies significantly, predisposing tested subjects to risks such as cerebral hypoxia and ischemia. Near-infrared spectroscopy (NIRS), a noninvasive technique performed for cerebral oximetry monitoring, was used in the present study to identify the minimum decrease in the end-tidal CO2 (ETCO2) during hyperventilation necessary to induce changes on NIRS. Materials and

Methods:

We recruited 46 volunteers with no preexisting medical conditions. Each subject was asked to breathe at a baseline rate (8−14 breaths/min) for 2 min and then to hyperventilate at a double respiratory rate for the next 4 min. The parameters recorded during the procedure were the regional cerebral oxyhemoglobin and deoxyhemoglobin concentrations via NIRS, ETCO2, and the respiratory rate.

Results:

During hyperventilation, ETCO2 values dropped (31.4 ± 12.2%) vs. baseline in all subjects. Changes in cerebral oximetry were observed only in those subjects (n = 30) who registered a decrease (%) in ETCO2 of 37.58 ± 10.34%, but not in the subjects (n = 16) for which the decrease in ETCO2 was 20.31 ± 5.6%. According to AUC-ROC analysis, a cutoff value of ETCO2 decrease >26% was found to predict changes in oximetry (AUC-ROC = 0.93, p < 0.0001). Seven subjects reported symptoms, such as dizziness, vertigo, and numbness, throughout the procedure.

Conclusions:

The rise in the respiratory rate alone cannot effectively predict the occurrence of a cerebral vasoconstrictor response induced by hyperventilation, and synchronous ETCO2 and cerebral oximetry monitoring could be used to validate this clinical test. NIRS seems to be a useful tool in predicting vasoconstriction following hyperventilation.