Your browser doesn't support javascript.
loading
On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beams.
Khan, Ahtesham Ullah; Nelson, Nicholas P; Culberson, Wesley S; DeWerd, Larry A.
Afiliação
  • Khan AU; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Nelson NP; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Culberson WS; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • DeWerd LA; Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Med Phys ; 50(2): 1105-1120, 2023 Feb.
Article em En | MEDLINE | ID: mdl-36334024
ABSTRACT

BACKGROUND:

In a recent study, we reported beam quality correction factors, fQ , in carbon ion beams using Monte Carlo (MC) methods for a cylindrical and a parallel-plate ionization chamber (IC). A non-negligible perturbation effect was observed; however, the magnitude of the perturbation correction due to the specific IC subcomponents was not included. Furthermore, the stopping power data presented in the International Commission on Radiation Units and Measurements (ICRU) report 73 were used, whereas the latest stopping power data have been reported in the ICRU report 90.

PURPOSE:

The aim of this study was to extend our previous work by computing fQ correction factors using the ICRU 90 stopping power data and by reporting IC-specific perturbation correction factors. Possible energy or linear energy transfer (LET) dependence of the fQ correction factor was investigated by simulating both pristine beams and spread-out Bragg peaks (SOBPs).

METHODS:

The TOol for PArticle Simulation (TOPAS)/GEANT4 MC code was used in this study. A 30 × 30 × 50 cm3 water phantom was simulated with a uniform 10 × 10 cm2 parallel beam incident on the surface. A Farmer-type cylindrical IC (Exradin A12) and two parallel-plate ICs (Exradin P11 and A11) were simulated in TOPAS using the manufacturer-provided geometrical drawings. The fQ correction factor was calculated in pristine carbon ion beams in the 150-450 MeV/u energy range at 2 cm depth and in the middle of the flat region of four SOBPs. The kQ correction factor was calculated by simulating the fQo correction factor in a 60 Co beam at 5 cm depth. The perturbation correction factors due to the presence of the individual IC subcomponents, such as the displacement effect in the air cavity, collecting electrode, chamber wall, and chamber stem, were calculated at 2 cm depth for monoenergetic beams only. Additionally, the mean dose-averaged and track-averaged LET was calculated at the depths at which the fQ was calculated.

RESULTS:

The ICRU 90 fQ correction factors were reported. The pdis correction factor was found to be significant for the cylindrical IC with magnitudes up to 1.70%. The individual perturbation corrections for the parallel-plate ICs were <1.0% except for the A11 pcel correction at the lowest energy. The fQ correction for the P11 IC exhibited an energy dependence of >1.00% and displayed differences up to 0.87% between pristine beams and SOBPs. Conversely, the fQ for A11 and A12 displayed a minimal energy dependence of <0.50%. The energy dependence was found to manifest in the LET dependence for the P11 IC. A statistically significant LET dependence was found only for the P11 IC in pristine beams only with a magnitude of <1.10%.

CONCLUSIONS:

The perturbation and kQ correction factor should be calculated for the specific IC to be used in carbon ion beam reference dosimetry as a function of beam quality.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Transferência Linear de Energia Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Transferência Linear de Energia Idioma: En Ano de publicação: 2023 Tipo de documento: Article