Your browser doesn't support javascript.
loading
Validation and method comparison for a point-of-care lateral flow assay measuring equine whole blood insulin concentrations.
Berryhill, Emily H; Urbina, Naomi S; Marton, Sam; Vernau, William; Alonso, Flavio H.
Afiliação
  • Berryhill EH; Departments of Medicine and Epidemiology, School of Veterinary Medicine, University of California-Davis, Davis, CA, USA.
  • Urbina NS; Fortis Life Sciences, San Diego, CA, USA.
  • Marton S; Fortis Life Sciences, San Diego, CA, USA.
  • Vernau W; Pathology, Microbiology, and Immunology, School of Veterinary Medicine, University of California-Davis, Davis, CA, USA.
  • Alonso FH; Pathology, Microbiology, and Immunology, School of Veterinary Medicine, University of California-Davis, Davis, CA, USA.
J Vet Diagn Invest ; 35(2): 124-131, 2023 Mar.
Article em En | MEDLINE | ID: mdl-36482705
ABSTRACT
The Wellness Ready Test (WRT) is a lateral flow, stall-side assay that measures equine insulin in whole blood and requires validation before recommending clinical use. We evaluated intra- and inter-assay precision and linearity and compared the WRT with a radioimmunoassay (RIA). Tested concentrations ranged from <139 to >695 pmol/L (<20 to >100 µIU/mL). For 20 replicates at each insulin level, intra-assay CVs of the WRT for insulin were 13.3%, 12.9%, and 15.3% at low (139-278 pmol/L; 20-40 µIU/mL), intermediate (278-417 pmol/L; 40-60 µIU/mL), and high (>417 pmol/L; >60 µIU/mL) concentrations, respectively. For 10 replicates at each level (3 assay lots), inter-assay CVs were 15.9%, 11.0%, and 11.7%, respectively. In the weighted linear regression of 5 measured insulin concentrations against expected concentrations, R2 = 0.98, slope = 1.02, and y-intercept = 14.4 pmol/L (2.08 µIU/mL). The Spearman correlation coefficient (rs) was 0.90 (95% CI 0.85-0.94) between the WRT and RIA; the WRT = f(RIA) Passing-Bablok regression yielded the fit, y = 1.005x + 24.3 pmol/L (3.50 µIU/mL). The WRT result averaged 10.4% higher than the RIA result, with targeted bias of 25.9, 26.1, and 26.7 pmol/L (3.74, 3.76, and 3.84 µIU/mL) for cutoffs used to diagnose insulin dysregulation of 312, 347, and 451 pmol/L (45, 50, and 65 µIU/mL). Assay clinical sensitivities, specificities, and accuracies determined at the 3 selected clinical cutoffs and using the RIA as gold standard were 87-95%, 92-96%, and 91-95%, respectively (n = 99 samples). Observed total error was 28.4-30.4%. The WRT had acceptable precision, excellent linearity, and good association with the RIA.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Insulina Limite: Animals Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Insulina Limite: Animals Idioma: En Ano de publicação: 2023 Tipo de documento: Article