Interfractional dose accumulation for MR-guided liver SBRT: Variation among algorithms is highly patient- and fraction-dependent.
Wahlstedt, Isak; George Smith, Abraham; Andersen, Claus Erik; Behrens, Claus Preibisch; Nørring Bekke, Susanne; Boye, Kristian; van Overeem Felter, Mette; Josipovic, Mirjana; Petersen, Jens; Risumlund, Signe Lenora; Tascón-Vidarte, José David; van Timmeren, Janita Elizabeth; Vogelius, Ivan Richter.
Afiliação
  • Wahlstedt I; Department of Health Technology, Technical University of Denmark, Anker Engelunds Vej 1, Bygning 101A, 2800 Kongens Lyngby, Denmark; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark; Department
  • George Smith A; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Computer Science, University of Copenhagen, Universitetsparken 1, 2100 Copenhagen, Denmark.
  • Andersen CE; Department of Health Technology, Technical University of Denmark, Anker Engelunds Vej 1, Bygning 101A, 2800 Kongens Lyngby, Denmark.
  • Behrens CP; Department of Health Technology, Technical University of Denmark, Anker Engelunds Vej 1, Bygning 101A, 2800 Kongens Lyngby, Denmark; Department of Oncology, Copenhagen University Hospital - Herlev and Gentofte (HGH), Borgmester Ib Juuls Vej 7, 2730 Herlev, Denmark.
  • Nørring Bekke S; Department of Oncology, Copenhagen University Hospital - Herlev and Gentofte (HGH), Borgmester Ib Juuls Vej 7, 2730 Herlev, Denmark.
  • Boye K; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark.
  • van Overeem Felter M; Department of Oncology, Copenhagen University Hospital - Herlev and Gentofte (HGH), Borgmester Ib Juuls Vej 7, 2730 Herlev, Denmark.
  • Josipovic M; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Health and Medical Sciences, University of Copenhagen, Blegdamsvej 3B, 2200 Copenhagen, Denmark.
  • Petersen J; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Computer Science, University of Copenhagen, Universitetsparken 1, 2100 Copenhagen, Denmark.
  • Risumlund SL; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark.
  • Tascón-Vidarte JD; Department of Computer Science, University of Copenhagen, Universitetsparken 1, 2100 Copenhagen, Denmark.
  • van Timmeren JE; Department of Radiation Oncology, University Hospital Zürich and University of Zürich, Zürich, Switzerland.
  • Vogelius IR; Department of Oncology, Centre for Cancer and Organ Diseases, Copenhagen University Hospital - Rigshospitalet (RH), Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Health and Medical Sciences, University of Copenhagen, Blegdamsvej 3B, 2200 Copenhagen, Denmark.
Radiother Oncol ; 182: 109448, 2023 05.
Article em En | MEDLINE | ID: mdl-36566988
ABSTRACT
BACKGROUND AND

PURPOSE:

Daily plan adaptations could take the dose delivered in previous fractions into account. Due to high dose delivered per fraction, low number of fractions, steep dose gradients, and large interfractional organ deformations, this might be particularly important for liver SBRT. This study investigates inter-algorithm variation of interfractional dose accumulation for MR-guided liver SBRT. MATERIALS AND

METHODS:

We assessed 27 consecutive MR-guided liver SBRT treatments of 67.5 Gy in three (n = 15) or 50 Gy in five fractions (n = 12), both prescribed to the GTV. We calculated fraction doses on daily patient anatomy, warped these doses to the simulation MRI using seven different algorithms, and accumulated the warped doses. Thus, we obtained differences in planned doses and warped or accumulated doses for each algorithm. This enabled us to calculate the inter-algorithm variations in warped doses per fraction and in accumulated doses per treatment course.

RESULTS:

The four intensity-based algorithms were more consistent with planned PTV dose than affine or contour-based algorithms. The mean (range) variation of the dose difference for PTV D95% due to dose warping by these intensity-based algorithms was 10.4 percentage points (0.3 to 43.7) between fractions and 8.6 (0.3 to 24.9) between accumulated treatment doses. As seen by these ranges, the variation was very dependent on the patient and the fraction being analyzed. Nevertheless, no correlations between patient or plan characteristics on the one hand and inter-algorithm dose warping variation on the other hand was found.

CONCLUSION:

Inter-algorithm dose accumulation variation is highly patient- and fraction-dependent for MR-guided liver SBRT. We advise against trusting a single algorithm for dose accumulation in liver SBRT.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Radioterapia de Intensidade Modulada Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiocirurgia / Radioterapia de Intensidade Modulada Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article