Your browser doesn't support javascript.
loading
Portable light-sheet optofluidic microscopy for 3D fluorescence imaging flow cytometry.
Son, Jeonghwan; Mandracchia, Biagio; Silva Trenkle, Aaron D; Kwong, Gabriel A; Jia, Shu.
Afiliação
  • Son J; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA 30332, USA. shu.jia@gatech.edu.
  • Mandracchia B; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA 30332, USA. shu.jia@gatech.edu.
  • Silva Trenkle AD; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA 30332, USA. shu.jia@gatech.edu.
  • Kwong GA; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA 30332, USA. shu.jia@gatech.edu.
  • Jia S; Parker H. Petit Institute of Bioengineering and Bioscience, Georgia Institute of Technology, Atlanta, GA 30332, USA.
Lab Chip ; 23(4): 624-630, 2023 02 14.
Article em En | MEDLINE | ID: mdl-36633262
ABSTRACT
Imaging flow cytometry (IFC) combines conventional flow cytometry with optical microscopy, allowing for high-throughput, multi-parameter screening of single-cell specimens with morphological and spatial information. However, current 3D IFC systems are limited by instrumental complexity and incompatibility with available microfluidic devices or operations. Here, we report portable light-sheet optofluidic microscopy (PLSOM) for 3D fluorescence cytometric imaging. PLSOM exploits a compact, open-top light-sheet configuration compatible with commonly adopted microfluidic chips. The system offers a subcellular resolution (2-4 µm) in all three dimensions, high throughput (∼1000 cells per s), and portability (30 cm (l) × 10 cm (w) × 26 cm (h)). We demonstrated PLSOM for 3D IFC using various phantom and cell systems. The low-cost and custom-built architecture of PLSOM permits easy adaptability and dissemination for broad 3D flow cytometric investigations.