Assessment of ultrasound shear wave elastography: An animal ex-vivo study.
Wang, Xiuming; Zhu, Jiaan; Liu, Yiqun; Li, Wenxue; Chen, Si; Zhang, Huabin.
Afiliação
  • Wang X; Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.
  • Zhu J; Department of Ultrasound, Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, Beijing, People's Republic of China.
  • Liu Y; Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.
  • Li W; Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.
  • Chen S; Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.
  • Zhang H; Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.
J Appl Clin Med Phys ; 24(4): e13924, 2023 Apr.
Article em En | MEDLINE | ID: mdl-36729737
OBJECTIVES: To explore the influence of the surrounding environment of the target tissue, lesion size, and rectangular sampling box size on shear wave speed (SWS). METHODS: The tendon SWS was acquired ex-vivo. Then the tendons were dissected and buried in the couplant (gel) and evaluated by two-dimensional shear wave elastography (2D-SWE). Finally, the tendons were placed in the isolated muscles to simulate the intramuscular lesions, and their elasticity was tested under two rectangular sampling box conditions. The isolated complete liver SWS was acquired. Similarly, the large and small pieces of livers were cut out, placed in the muscles, and assessed by SWE under two rectangular sampling box conditions. The SWS acquired under different conditions was compared. Variability was evaluated using the coefficient of variation (CV). The intraclass correlation coefficient (ICC) was used to evaluate repeatability. RESULTS: The SWS of the tendons ex-vivo, buried in the couplant and placed in the isolated muscles showed significant differences (p < 0.001). The ex-vivo condition produced the highest SWS and CV values. There were significant differences in SWS of livers with different sizes placed in muscles (p < 0.001). The highest SWS value was associated with small pieces of livers. No significant difference was found in SWS acquired under different rectangular box sizes (p > 0.05). CONCLUSIONS: Under the present study conditions, the surrounding environment of the target tissue makes a big difference to lesion SWS values. The lesion size will affect the assessment of its inherent elasticity. The size of the sampling frame has no significant effect on the tissue SWS.