A prospective survey of veterinary anesthesia equipment in Alberta, Canada, using a standardized checkout procedure.
Marchiori, Jocelyn J M; Prebble, Melanie J; Pang, Daniel S J.
Afiliação
  • Marchiori JJM; Western Veterinary Specialist and Emergency Centre, 1802 10th Avenue SW, Calgary, Alberta T3C 0J8 (Marchiori); Dispomed Ltd, 745 Rue Nazaire Laurin, Joliette, Québec J6E 0L6 (Prebble); Department of Veterinary Clinical and Diagnostic Sciences, Faculty of Veterinary Medicine, University of Calgary, 3280 Hospital Drive NW, Calgary, Alberta T2N 4Z6 (Pang); Department of Clinical Sciences, Faculty of Veterinary Medicine, Université de Montréal, 3200, rue Sicotte Saint-Hyacinthe, Québec J2S 2M2 (Pang
  • Prebble MJ; Western Veterinary Specialist and Emergency Centre, 1802 10th Avenue SW, Calgary, Alberta T3C 0J8 (Marchiori); Dispomed Ltd, 745 Rue Nazaire Laurin, Joliette, Québec J6E 0L6 (Prebble); Department of Veterinary Clinical and Diagnostic Sciences, Faculty of Veterinary Medicine, University of Calgary, 3280 Hospital Drive NW, Calgary, Alberta T2N 4Z6 (Pang); Department of Clinical Sciences, Faculty of Veterinary Medicine, Université de Montréal, 3200, rue Sicotte Saint-Hyacinthe, Québec J2S 2M2 (Pang
  • Pang DSJ; Western Veterinary Specialist and Emergency Centre, 1802 10th Avenue SW, Calgary, Alberta T3C 0J8 (Marchiori); Dispomed Ltd, 745 Rue Nazaire Laurin, Joliette, Québec J6E 0L6 (Prebble); Department of Veterinary Clinical and Diagnostic Sciences, Faculty of Veterinary Medicine, University of Calgary, 3280 Hospital Drive NW, Calgary, Alberta T2N 4Z6 (Pang); Department of Clinical Sciences, Faculty of Veterinary Medicine, Université de Montréal, 3200, rue Sicotte Saint-Hyacinthe, Québec J2S 2M2 (Pang
Can Vet J ; 64(2): 159-166, 2023 02.
Article em En | MEDLINE | ID: mdl-36733657
ABSTRACT

Background:

In both human and veterinary medicine, it is recommended that an anesthetic machine checkout procedure (preuse check) be performed daily, with some items tested before each case, to confirm safe function and the check results recorded.

Objective:

The objective of this prospective study was to evaluate anesthetic machines in private veterinary clinics in Alberta (Canada) using a standardized checkout procedure. Animals and procedures One-hundred consecutive anesthetic machines were assessed. For each item of the checkout procedure, a "pass," "fail," or "not applicable" score was awarded. "Not applicable" indicated an item that could not be evaluated.

Results:

Few machines (10%) evaluated had a secondary oxygen supply, no machines had an oxygen supply pressure alarm, and leaks were identified in 31 and 17% of rebreathing and non-rebreathing systems, respectively. Thirty-nine percent of machines did not have a high-pressure circuit alarm, 86% of machines were attached to an active scavenging system, although it was improperly connected in 56% of cases, and only 2% of machines were accompanied by a checkout log. Conclusion and clinical relevance There was widespread variation in anesthetic machine standards and function, highlighting the value of performing a regular machine checkout procedure in creating a foundation for safe anesthetic practice.
Une enquête prospective sur l'équipement d'anesthésie vétérinaire en Alberta, au Canada, à l'aide d'une procédure de vérification normalisée. Historique En médecine humaine et vétérinaire, il est recommandé qu'une procédure de vérification de la machine d'anesthésie (vérification avant utilisation) soit effectuée quotidiennement, avec certains éléments testés avant chaque cas, pour confirmer le fonctionnement sécuritaire et les résultats de la vérification enregistrés. Objectif L'objectif de cette étude prospective était d'évaluer les machines d'anesthésie dans les cliniques vétérinaires privées de l'Alberta (Canada) en utilisant une procédure de contrôle standardisée. Animaux et procédures Cent machines d'anesthésie consécutives ont été évaluées. Pour chaque élément de la procédure de vérification, une note « réussite ¼, « échec ¼ ou « non applicable ¼ a été attribuée. « Sans objet ¼ indique un élément qui n'a pas pu être évalué. Résultats Peu de machines (10 %) évaluées avaient une alimentation secondaire en oxygène, aucune machine n'avait d'alarme de pression d'alimentation en oxygène, et des fuites ont été identifiées dans 31 et 17 % des systèmes de ré-inhalation et sans ré-inhalation, respectivement. Trente-neuf pourcents des machines n'avaient pas d'alarme de circuit haute pression, 86 % des machines étaient rattachées à un système de balayage actif, bien qu'il soit mal raccordé dans 56 % des cas, et seulement 2 % des machines étaient accompagnées d'un journal de sortie. Conclusion et pertinence clinique Il y avait une grande variation dans les normes et le fonctionnement des machines d'anesthésie, soulignant l'importance d'effectuer une procédure de vérification régulière de la machine pour créer une base pour une pratique anesthésique sûre.(Traduit par Dr Serge Messier).