Quantitation of neurofilament light chain protein in serum and cerebrospinal fluid from patients with multiple sclerosis using the MSD R-PLEX NfL assay.
Ulndreaj, Antigona; Sohaei, Dorsa; Thebault, Simon; Pons-Belda, Oscar D; Fernandez-Uriarte, Amaia; Campbell, Christopher; Cheo, David; Stengelin, Martin; Sigal, George; Freedman, Mark S; Scarisbrick, Isobel A; Prassas, Ioannis; Diamandis, Eleftherios P.
Afiliação
  • Ulndreaj A; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Toronto, ON, Canada.
  • Sohaei D; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Toronto, ON, Canada.
  • Thebault S; Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
  • Pons-Belda OD; Department of Medicine, Univeristy of Ottawa, The Ottawa Hospital Research Institute, Ottawa, ON, Canada.
  • Fernandez-Uriarte A; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Toronto, ON, Canada.
  • Campbell C; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Toronto, ON, Canada.
  • Cheo D; Meso Scale Diagnostics, LLC. (MSD), Rockville, MD, USA.
  • Stengelin M; Meso Scale Diagnostics, LLC. (MSD), Rockville, MD, USA.
  • Sigal G; Meso Scale Diagnostics, LLC. (MSD), Rockville, MD, USA.
  • Freedman MS; Meso Scale Diagnostics, LLC. (MSD), Rockville, MD, USA.
  • Scarisbrick IA; Department of Medicine, Univeristy of Ottawa, The Ottawa Hospital Research Institute, Ottawa, ON, Canada.
  • Prassas I; Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, USA.
  • Diamandis EP; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Toronto, ON, Canada.
Diagnosis (Berl) ; 10(3): 275-280, 2023 08 01.
Article em En | MEDLINE | ID: mdl-36788117
ABSTRACT

OBJECTIVES:

Neurofilament light (NfL) chain is a marker of neuroaxonal damage in various neurological diseases. Here we quantitated NfL levels in the cerebrospinal fluid (CSF) and serum from patients with multiple sclerosis (MS) and controls, using the R-PLEX NfL assay, which employs advanced Meso Scale Discovery® (MSD) electrochemiluminescence (ECL)-based detection technology.

METHODS:

NfL was quantitated in samples from 116 individuals from two sites (Ottawa Hospital Research Institute and Mayo Clinic), consisting of patients with MS (n=71) and age- and sex-matched inflammatory neurological controls (n=13) and non-inflammatory controls (n=32). Correlation of NfL levels between CSF and serum was assessed in paired samples in a subset of MS patients and controls (n=61). Additionally, we assessed the correlation between NfL levels obtained with MSD's R-PLEX® and Quanterix's single molecule array (Simoa®) assays in CSF and serum (n=32).

RESULTS:

Using the R-PLEX, NfL was quantitated in 99% of the samples tested, and showed a broad range in the CSF (82-500,000 ng/L) and serum (8.84-2,014 ng/L). Nf-L levels in both biofluids correlated strongly (r=0.81, p<0.0001). Lastly, Nf-L measured by MSD's R-PLEX and Quanterix's Simoa assays were highly correlated for both biofluids (CSF r=0.94, p<0.0001; serum r=0.95, p<0.0001).

CONCLUSIONS:

We show that MSD's R-PLEX NfL assay can reliably quantitate levels of NfL in the CSF and serum from patients with MS and controls, where levels correlate strongly with Simoa.