Your browser doesn't support javascript.
loading
Determination of a Serum 25-Hydroxyvitamin D Reference Ranges in Japanese Adults Using Fully Automated Liquid Chromatography-Tandem Mass Spectrometry.
Miyamoto, Hiroyasu; Kawakami, Daisuke; Hanafusa, Nobuhiro; Nakanishi, Tsuyoshi; Miyasaka, Masaki; Furutani, Yutaka; Ikeda, Yuichi; Ito, Kyoko; Kato, Tomohiro; Yokoyama, Keitaro; Arakawa, Shoutaro; Saito, Mitsuru; Furusho, Tadasu; Matsuura, Tomokazu; Ochi, Sae.
Afiliação
  • Miyamoto H; Department of Laboratory Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan; Department of Central Clinical Laboratory, The Jikei University Hospital, Minato City, Tokyo, Japan; Department of International Food and Agricultural Science, Faculty of International Agriculture
  • Kawakami D; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto city, Kyoto, Japan; Analytical Business Unit, Shimadzu Europa GmbH, Duisburg, Germany.
  • Hanafusa N; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto city, Kyoto, Japan.
  • Nakanishi T; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto city, Kyoto, Japan.
  • Miyasaka M; Department of Laboratory Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan.
  • Furutani Y; Department of Laboratory Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan.
  • Ikeda Y; Department of Central Clinical Laboratory, The Jikei University Hospital, Minato City, Tokyo, Japan.
  • Ito K; Centre for Preventive Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan.
  • Kato T; Centre for Preventive Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan.
  • Yokoyama K; Harumi Toriton Clinic of The Jikei University Hospital, Chuo City, Tokyo, Japan; Department of Health Science, The Graduate School, The Jikei University School of Medicine, Tokyo, Japan.
  • Arakawa S; Department of Orthopedics, The Jikei University School of Medicine, Minato-ku, Tokyo, Japan.
  • Saito M; Department of Orthopedics, The Jikei University School of Medicine, Minato-ku, Tokyo, Japan.
  • Furusho T; Department of International Food and Agricultural Science, Faculty of International Agriculture and Food Studies, Tokyo University of Agriculture, Setagaya City, Tokyo, Japan.
  • Matsuura T; Department of Laboratory Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan.
  • Ochi S; Department of Laboratory Medicine, The Jikei University School of Medicine, Minato City, Tokyo, Japan. Electronic address: ochisae1024@jikei.ac.jp.
J Nutr ; 153(4): 1253-1264, 2023 04.
Article em En | MEDLINE | ID: mdl-36806449
ABSTRACT

BACKGROUND:

Despite an increasing interest in vitamin D status, a reference range of the nutrient has not been fully established. This is partly due to a paucity of standardized measuring systems with high throughput. In addition, the range may vary by populations and may change with modernization of lifestyles.

OBJECTIVES:

This study aims to calculate the current reference concentration of 25-hydroxyvitamin D (25(OH)D) among healthy people living in an urban area in Japan.

METHODS:

A newly developed fully automated liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) system was used to measure serum 25(OH)D concentrations. Reproducibility was assessed by measuring standardized samples. Accuracy was validated by comparing with commercially available immunoassays. Then, mass screening was conducted targeting participants who received medical checkups in Tokyo from April 2019 to March 2020, and the reference ranges were calculated.

RESULTS:

The coefficients of variations of interoperator and interday reproducibility were 4.1%-8.5% and 3.7%-8.0% for 25-hydroxyvitamin D2 (25(OH)D2) and 4.7%-7.0% and 4.0%-6.9% for 25-hydroxyvitamine D3, respectively. The measured total 25(OH)D concentrations correlated well with those measured by immunoassays. In total, 5518 participants were measured for 25(OH)D concentrations, among whom 98% showed inadequate concentrations (<30 ng/mL). The reference ranges of total 25(OH)D for female, male, and total participants were 7-30 ng/mL, 5-27 ng/mL, and 6-29 ng/mL, respectively. After excluding those with abnormal renal and liver function, the range was 6-30 ng/mL.

CONCLUSIONS:

The high prevalence of vitamin D insufficiency among seemingly healthy population may be attributed to lifestyle characteristics of people living in urban areas of Japan, including spending less time outdoors and lower intake of traditional foods. Longitudinal follow-up and mass screenings targeting different population will help elucidate reasons for discrepancies between official guidelines and the observed concentrations, to which the well-validated measurement system is essential.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vitamina D / Deficiência de Vitamina D / Cromatografia Líquida / Espectrometria de Massas em Tandem / População do Leste Asiático Tipo de estudo: Diagnostic_studies / Etiology_studies / Risk_factors_studies / Screening_studies Limite: Adult / Female / Humans / Male País/Região como assunto: Asia Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Vitamina D / Deficiência de Vitamina D / Cromatografia Líquida / Espectrometria de Massas em Tandem / População do Leste Asiático Tipo de estudo: Diagnostic_studies / Etiology_studies / Risk_factors_studies / Screening_studies Limite: Adult / Female / Humans / Male País/Região como assunto: Asia Idioma: En Ano de publicação: 2023 Tipo de documento: Article