Your browser doesn't support javascript.
loading
Dose uncertainty due to needle-tip localization error in prostate seed implantation.
Xu, Zhengzheng; Arsenault, Theodore H; Traughber, Bryan; Ove, Roger; Podder, Tarun K.
Afiliação
  • Xu Z; Department of Radiation Oncology, Seidman Cancer Center, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, Cleveland, OH 44106, USA.
  • Arsenault TH; Department of Radiation Oncology, University of Southern California, Norris Cancer Hospital, Los Angeles, CA 90033, USA.
  • Traughber B; Department of Radiation Oncology, Seidman Cancer Center, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, Cleveland, OH 44106, USA.
  • Ove R; Department of Radiation Oncology, Pennsylvania State University, Hershey, PA 17033, USA.
  • Podder TK; Department of Radiation Oncology, Seidman Cancer Center, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, Cleveland, OH 44106, USA.
J Contemp Brachytherapy ; 14(6): 582-589, 2022 Dec.
Article em En | MEDLINE | ID: mdl-36819470
ABSTRACT

Purpose:

This study quantified the dosimetric uncertainty caused by needle-tip detection errors in ultrasound images due to bevel-tip orientation differences, with respect to the location on template grid. Material and

methods:

Trans-rectal ultrasound (TRUS) system with physical template grid and 18-gauge bevel-tip brachytherapy needles were used. TRUS was set at 6.5 MHz in water phantom, and measurements were taken with 50% and 100% B-mode TRUS gains. Needle-tip localization errors were then retrospectively applied back to 45 prostate seed implant plans to evaluate the important planning parameters for the prostate (D90, V100, V150, and V200), urethra (D10 and D30), and rectum (V100, D2cc, and D0.1cc), following the ABS and AAPM TG-137 guidelines.

Results:

The needle-tip detection errors for 50% and 100% TRUS gains were 3.7 mm (max) and 5.2 mm (max), respectively. The observed significant decrease in prostate coverage (mean D90 lower by 12.8%, and V100 lower by 3.9% for smaller prostates) after seed placements were corrected by compensating the needle-tip detection errors. Apex of the prostate was hotter, and the base was cooler. Dosimetric difference for urethral and rectal parameters were not statistically significant.

Conclusions:

This study revealed that the beveled needle-tip orientation could considerably impact the needle tips detection accuracy, based on which the seeds might be delivered. These errors can lead to significant dosimetric uncertainty in prostate seed implantation.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2022 Tipo de documento: Article