Your browser doesn't support javascript.
loading
Exploring the Radioprotective Indium (III) Oxide Screens for Mammography Scans Using a Three-Layer Heterogeneous Breast Phantom and MCNPX: A Comparative Study Using Clinical Findings.
ALMisned, Ghada; Elshami, Wiam; Kilic, Gokhan; Ilik, Erkan; Rabaa, Elaf; Zakaly, Hesham M H; Ene, Antoaneta; Tekin, Huseyin O.
Afiliação
  • ALMisned G; Department of Physics, College of Science, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
  • Elshami W; Medical Diagnostic Imaging Department, College of Health Sciences, University of Sharjah, Sharjah 27272, United Arab Emirates.
  • Kilic G; Department of Physics, Faculty of Science, Eskisehir Osmangazi University, Eskisehir 26040, Türkiye.
  • Ilik E; Department of Physics, Faculty of Science, Eskisehir Osmangazi University, Eskisehir 26040, Türkiye.
  • Rabaa E; Medical Diagnostic Imaging Department, College of Health Sciences, University of Sharjah, Sharjah 27272, United Arab Emirates.
  • Zakaly HMH; Institute of Physics and Technology, Ural Federal University, 620002 Ekaterinburg, Russia.
  • Ene A; Physics Department, Faculty of Science, Al-Azhar University, Assiut 71524, Egypt.
  • Tekin HO; INPOLDE Research Center, Department of Chemistry, Physics and Environment, Faculty of Sciences and Environment, Dunarea de Jos University of Galati, 47 Domneasca Street, 800008 Galati, Romania.
Medicina (Kaunas) ; 59(2)2023 Feb 09.
Article em En | MEDLINE | ID: mdl-36837529
ABSTRACT

Background:

During mammography, a lead-acrylic protective screen is recommended to reduce radiation exposure to the unexposed breast.

Objectives:

This research study aimed to construct an Indium-(III)-oxide-rich tellurite-glass screen (TZI8) and compare its performance to that of lead acrylic. Materials and

Methods:

A three-layer heterogeneous-breast phantom was developed, using the MCNPX (version 2.7.0) Monte Carlo code. An MCNPX-simulation geometry was designed and implemented, using the lead-acrylic and TZI8 shielding screens between the right and left breast. Next, the reliability of the phantom and the variations in absorption between the lead-acrylic and TZI8 glass were investigated.

Results:

The findings show that the TZI8-protective-glass screen offers significantly greater radioprotection than the lead-acrylic material. The quantity of total dose absorbed in the unexposed breast was much lower for TZI8 than for lead-based acrylic. The TZI8-glass screen gives about 60% more radioprotection than the lead-acrylic screen.

Conclusion:

Considering the toxic lead in the structure that may be hazardous to the human tissues, the TZI8-glass screen may be used in mammography examination to provide greater radioprotection than the lead-acrylic screen, in order to greatly reduce the dose to the unexposed breast.