Two-photon absorption spectrometers for near infrared.
Rev Sci Instrum
; 94(5)2023 May 01.
Article
em En
| MEDLINE
| ID: mdl-37140339
We present a Silicon-based Charge-Coupled Device (Si-CCD) sensor applied as a cost-effective spectrometer for femtosecond pulse characterization in the Near Infrared region in two different configurations: two-Fourier and Czerny-Turner setups. To test the spectrometer's performance, a femtosecond Optical Parametric Oscillator with a tuning range between 1100 and 1700 nm and a femtosecond Erbium-Doped Fiber Amplifier at 1582 nm were employed. The nonlinear spectrometer operation is based on the Two-Photon Absorption effect generated in the Si-CCD sensor. The achieved spectrometer resolution was 0.6 ± 0.1 nm with a threshold peak intensity of 2×106Wcm2. An analysis of the nonlinear response as a function of the wavelength, the response saturation, and the criteria to prevent it are also presented.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article