ABSTRACT

PURPOSE:

This paper aims to investigate the impact of the channel numbers on the performance of B1+ mapping, by using the Bloch-Siegert shift (BSS) method. B1+ mapping plays a crucial role in various brain imaging protocols. THEORY AND

METHODS:

We simulated the radiofrequency field of the human head model in six groups of multi-channel receive coil with a range of different channel numbers. MR signals were synthesized according to the standard BSS sequence, with quantified Gaussian added. Next, we combined the signals of each channel to reconstruct the B1+ map by weighted averaging and maximum likelihood estimation strategies and evaluate the bias by relative standard deviation of each coil.

RESULTS:

The simulation results revealed that the accuracy of B1+ maps improved with the increasing of channel numbers, meanwhile the per channel efficiency of B1+maps accuracy gradually decrease. Both trends slowed down when the channel numbers reached 12 or above.

CONCLUSION:

Our finding suggests that increasing the channel numbers can improve the accuracy of B1+map. However, a diminishing efficiency of per channel accuracy improvement was overserved, indicating that the relationship between quality of B1+ map and the channel numbers is nonlinear. Based on these findings, our study provides a reference for determining channel numbers to achieve a balance of coil selection and manufacturing cost. It also provides a theoretical basis for evaluating other B1+ mapping techniques.