Your browser doesn't support javascript.
loading
Evaluation of a deuterated triarylmethyl spin probe for in vivo R2 ∗-based EPR oximetric imaging with enhanced dynamic range.
Kishimoto, Shun; Devasahayam, Nallathamby; Chandramouli, Gadisetti V R; Murugesan, Ramachandran; Otowa, Yasunori; Yamashita, Kota; Yamamoto, Kazutoshi; Brender, Jeffrey R; Krishna, Murali C.
Afiliação
  • Kishimoto S; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Devasahayam N; Urologic Oncology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Chandramouli GVR; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Murugesan R; GenEpria Consulting Inc., Columbia, Maryland, USA.
  • Otowa Y; Karpaga Vinayaga Institute of Medical Sciences and Research Center, Chengalpattu, Tamil Nadu, India.
  • Yamashita K; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Yamamoto K; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Brender JR; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
  • Krishna MC; Radiation Biology Branch, Center for Cancer Research, NCI, Bethesda, Maryland, USA.
Magn Reson Med ; 91(1): 413-423, 2024 01.
Article em En | MEDLINE | ID: mdl-37676121
ABSTRACT

PURPOSE:

In this study, we compared two triarylmethyl (TAM) spin probes, Ox071 and Ox063 for their efficacy in measuring tissue oxygen levels under hypoxic and normoxic conditions by R2 *-based EPR oximetry.

METHODS:

The R2 * dependencies on the spin probe concentration and oxygen level were calibrated using deoxygenated 1, 2, 5, and 10 mM standard solutions and 2 mM solutions saturated at 0%, 2%, 5%, 10%, and 21% of oxygen. For the hypoxic model, in vivo imaging of a MIA PaCa-2 tumor implanted in the hind leg of a mouse was performed on successive days by R2 *-based EPR oximetry using either Ox071 or Ox063. For the normoxic model, renal imaging of healthy athymic mice was performed using both spin probes. The 3D images were reconstructed by single point imaging and multi-gradient technique was used to determine R2 * maps.

RESULTS:

The signal intensities of Ox071 were approximately three times greater than that of Ox063 in the entire partial pressure of oxygen (pO2 ) range investigated. The histograms of the tumor pO2 images were skewed for both spin probes, and Ox071 showed more frequency counts at pO2 > 32 mm Hg. In the normoxic kidney model, there was a clear delineation between the high pO2 cortex and the low pO2 medulla regions. The histogram of high-resolution kidney oximetry image using Ox071 was nearly symmetrical and frequency counts were seen up to 55 mm Hg, which were missed in Ox063 imaging.

CONCLUSION:

As an oximetric probe, Ox071 has clear advantages over Ox063 in terms of sensitivity and the pO2 dynamic range.