Your browser doesn't support javascript.
loading
Practical guide for setting up a Fourier light-field microscope.
Appl Opt ; 62(16): 4228-4235, 2023 Jun 01.
Article em En | MEDLINE | ID: mdl-37706910
ABSTRACT
A practical guide for the easy implementation of a Fourier light-field microscope is reported. The Fourier light-field concept applied to microscopy allows the capture in real time of a series of 2D orthographic images of microscopic thick dynamic samples. Such perspective images contain spatial and angular information of the light-field emitted by the sample. A feature of this technology is the tight requirement of a double optical conjugation relationship, and also the requirement of NA matching. For these reasons, the Fourier light-field microscope being a non-complex optical system, a clear protocol on how to set up the optical elements accurately is needed. In this sense, this guide is aimed to simplify the implementation process, with an optical bench and off-the-shelf components. This will help the widespread use of this recent technology.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline Idioma: En Ano de publicação: 2023 Tipo de documento: Article