Your browser doesn't support javascript.
loading
Age- and sex-adjusted reference intervals for steroid hormones measured by liquid chromatography-tandem mass spectrometry using a widely available kit.
Kunz, Sonja; Wang, Xiao; Ferrari, Uta; Drey, Michael; Theodoropoulou, Marily; Schilbach, Katharina; Reincke, Martin; Heier, Margit; Peters, Annette; Koenig, Wolfgang; Zeller, Tanja; Thorand, Barbara; Bidlingmaier, Martin.
Afiliação
  • Kunz S; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Wang X; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Ferrari U; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Drey M; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Theodoropoulou M; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Schilbach K; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Reincke M; Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
  • Heier M; Institute of Epidemiology, Helmholtz Zentrum München, German Research Center for Environmental Health, Neuherberg, Germany.
  • Peters A; KORA Study Centre, University Hospital of Augsburg, Augsburg, Augsburg, Germany.
  • Koenig W; Institute of Epidemiology, Helmholtz Zentrum München, German Research Center for Environmental Health, Neuherberg, Germany.
  • Zeller T; German Centre for Cardiovascular Research (DZHK), Partner site Munich Heart Alliance, Munich, Germany.
  • Thorand B; German Centre for Cardiovascular Research (DZHK), Partner site Munich Heart Alliance, Munich, Germany.
  • Bidlingmaier M; German Heart Centre Munich, Technical University of Munich, Munich, Germany.
Endocr Connect ; 13(1)2024 Jan 01.
Article em En | MEDLINE | ID: mdl-37938144
ABSTRACT

Objective:

Measurements utilizing commercially available sets of reagents for determination of steroid hormone profiles by liquid chromatography-tandem mass spectrometry (LC-MS/MS) have become increasingly important for routine laboratories. However, method-specific publications of reference intervals obtained from sufficiently large studies are often missing.

Methods:

After validation of performance characteristics, a widely available kit for steroid analysis by LC-MS/MS was used to measure concentrations of 15 endogenous steroids (aldosterone, cortisol, cortisone, corticosterone, 11-deoxycortisol, 21-deoxycortisol, dehydroepiandrosterone sulfate, estradiol, testosterone, androstenedione, dihydrotestosterone, dehydroepiandrosterone, 17-hydroxyprogesterone, 11-deoxycorticosterone, progesterone) in more than 500 blood samples from a population-based study. While randomly selected from a larger cohort, the samples equally represented both sexes and covered a wide range of adult age groups. Age- and sex-specific reference intervals were calculated, and correlation with BMI was assessed.

Results:

Performance characteristics of the assay matched expectations for 9 of 15 steroids. For most of them, reference intervals obtained from our study population were comparable to those reported by others, with age and sex being the major determinants. A sex-specific correlation with BMI was found for seven steroids. We identified limitations regarding sensitivity of the method for quantification of progesterone in males and postmenopausal females. Concentrations of aldosterone, 21-deoxycortisol, estradiol, 11-deoxycorticosterone, and dihydrotestosterone could not be quantified in a large percentage of samples.

Conclusions:

The reference intervals for nine steroids will support meaningful interpretation for steroid profiles as measured by a widely used kit for LC-MS/MS-based quantification. Laboratories using such kits must be aware of potential limitations in sensitivity for some steroids included in the profile. Significance Statement Quantification of steroid hormones is a cornerstone for diagnosis of several diseases. Commonly used immunoassays have limitations in specificity. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a promising alternative, particularly if methods are harmonized across laboratories. The use of kits from commercial suppliers might support this. Clinical interpretation of steroid concentrations requires availability of appropriate reference intervals (RIs), but studies on RIs reported in the literature differ in preanalytical and analytical procedures. Here, we provide RIs for steroids measured by a widely available kit under preanalytical conditions mirroring common clinical practice. Such RIs might facilitate interpretation for those using the same method and comparable conditions in clinical routine.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article