A Y-shape connection device for pediatric patients with an indwelling catheter (Dia = 8Fr) during urodynamic studies, especially for filling phase measurements: a single-center prospective study for safety and effectiveness.
Yang, Shi-Qin; Zeng, Xiao; Peng, Liao; Shen, Si-Hong; Zhang, Jie; Huang, Zhi-Hui; Shen, Hong; Luo, De-Yi; Jin, Tao.
Afiliação
  • Yang SQ; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Zeng X; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Peng L; Pelvic Floor Diseases Center, West China Tianfu Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Shen SH; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Zhang J; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Huang ZH; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Shen H; Pelvic Floor Diseases Center, West China Tianfu Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Luo DY; Operating Room, West China Hospital, Sichuan University/West China School of Nursing, Sichuan University, Chengdu, Sichuan, China.
  • Jin T; Department of Urology, Institute of Urology, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
Front Pediatr ; 11: 1271417, 2023.
Article em En | MEDLINE | ID: mdl-38027283
Introduction: This prospective study aimed to assess the effectiveness of a Y-shape connection device in reducing pain and bleeding in pediatric patients with indwelling catheters during urodynamic studies (UDS), while also obtaining effective results in the filling phase. Methods: A total of 45 pediatric patients with a mean age of 13 years were included, all of whom underwent both a UDS with the Y-shape connection device (Method A) and a standard UDS procedure (Method B). Results: The Y-shape connection device demonstrated similar overall urodynamic parameters compared to the standard procedure, while also resulting in significantly less bleeding (P = 0.006) and lower VAS scores during (1.12 ± 0.58 vs. 3.88 ± 1.01, P = 0.001) and after (0.12 ± 0.08 vs 2.91 ± 0.89, P = 0.001) the procedure. No adverse events were reported at the 1-month follow-up. Discussion: These findings suggest that the Y-shape connection device can effectively reduce pain and bleeding during and after UDS in pediatric patients with indwelling catheters (Dia = 8Fr), while also obtaining effective results in the filling phase. Therefore, this Y-shape connection device has a more significant value for children who require urodynamic studies and place more emphasis on filling phase parameters. Clinical trial registration: ChiCTR2300068280.