Clinical specificity of two assays for immunoglobulin kappa and lambda free light chains.
Farnsworth, Christopher W; Roemmich, Brittany; Spears, Grant M; Murray, David L; Dispenzieri, Angela; Willrich, Maria Alice V.
Afiliação
  • Farnsworth CW; Washington University in St. Louis, St. Louis, MO, USA.
  • Roemmich B; Washington University in St. Louis, St. Louis, MO, USA.
  • Spears GM; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Murray DL; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Dispenzieri A; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Willrich MAV; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
Clin Chem Lab Med ; 62(5): 929-938, 2024 Apr 25.
Article em En | MEDLINE | ID: mdl-38044587
ABSTRACT

OBJECTIVES:

Free light chain (FLC) assays and the ratio of κ/λ are recommended for diagnosis, prognosis and monitoring of plasma cell dyscrasias (PCD). Limited data exists on FLC clinical specificity in patients diagnosed with other conditions.

METHODS:

We assessed the κ, λ, and κ/λ FLC ratio using the FreeLite assay and the Sebia FLC ELISA assay in 176 patients with clinical presentations of fatigue, anemia, polyclonal hypergammaglobulinemia, joint disorders, kidney disease and non PCD-cancers with no monoclonal protein observed on serum protein electrophoresis or MASS-FIX immunoglobulin isotyping. Manufacturer defined reference intervals (RI) and glomerular filtration rate (GFR) specific RI (renal RI) were utilized.

RESULTS:

For the κ/λ ratio, 68.7 % (121/176) of specimens on the FreeLite and 87.5 % (154/176) of specimens on the Sebia assay were within RI. For κ, 68.2 % (120/176) and 72.2 % (127/176) of results were outside RI for FreeLite and Sebia respectively. For λ, 37.5 % (66/176) and 84.1 % (148/176) of FreeLite and Sebia results were outside RI. With FreeLite and Sebia, patients with kidney disease (n=25) had the highest κ/λ ratios. 44 patients (25.0 %) had GFR <60 mL/min/BSA. When renal RI were applied, 13.6 % had a FLCr outside the renal RI with FreeLite, and 4.5 % with Sebia.

CONCLUSIONS:

In a cohort of patients with signs and symptoms suggestive of PCDs, but ultimately diagnosed with other conditions, Sebia FLC had improved clinical specificity relative to FreeLite, if one was using an abnormal κ/λ ratio as a surrogate for monoclonality.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Paraproteinemias / Nefropatias Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Paraproteinemias / Nefropatias Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article