Validation of the DBP-1333b upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020).
Blood Press Monit
; 29(3): 149-155, 2024 Jun 01.
Article
em En
| MEDLINE
| ID: mdl-38193368
ABSTRACT
To evaluate the accuracy of the DBP-1333b upper-arm blood pressure (BP) measuring device in the adult population according to the AAMI/ESH/ISO universal standard (ISO 81060-22018+Amd.12020). Subjects were recruited in the adult population. The test device was an arm-type electronic sphygmomanometer (DBP-1333b) and the reference device was a desktop sphygmomanometer (XJ11D). Using the BP data measured by the desktop sphygmomanometer as reference BP, the accuracy of the non-invasive BP module of the test device was evaluated to determine whether it met the requirements. Data from 90 individuals were analysed. According to Criterion 1, the mean difference of SBP between the test and reference device was 0.19â
mmHg and the SD was 7.45â
mmHg. The mean difference of DBP was -0.59â
mmHg and the SD was 6.47â
mmHg. The mean difference of both SBP and DBP was less than 5â
mmHg, and the SD was less than 8â
mmHg, which met the requirements. According to Criterion 2, SD of SBP was 5.79â
mmHg, which was less than 6.95â
mmHg and met the requirements. The SD of DBP was 5.58â
mmHg, which was less than 6.93â
mmHg and met the requirements. It was concluded that the DBP-1333b complies with the AAMI/ESH/ISO universal standard (ISO 81060-22018+Amd.12020) and can be recommended for use by the adults.