Effects of non-stationary blur on texture biomarkers of bone using Ultra-High Resolution CT.
Shi, G; Quevedo Gonzalez, F J; Breighner, R E; Carrino, J A; Siewerdsen, J H; Zbijewski, W.
Afiliação
  • Shi G; Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD USA 21205.
  • Quevedo Gonzalez FJ; Department of Biomechanics, Hospital for Special Surgery, New York, NY USA 10021.
  • Breighner RE; Department of Biomechanics, Hospital for Special Surgery, New York, NY USA 10021.
  • Carrino JA; Hospital for Special Surgery, Radiology & Imaging, New York, NY USA 10021.
  • Siewerdsen JH; MD Anderson Cancer Center, Houston, TX USA 77030.
  • Zbijewski W; Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD USA 21205.
Article em En | MEDLINE | ID: mdl-38226358
ABSTRACT

Purpose:

To advance the development of radiomic models of bone quality using the recently introduced Ultra-High Resolution CT (UHR CT), we investigate inter-scan reproducibility of trabecular bone texture features to spatially-variant azimuthal and radial blurs associated with focal spot elongation and gantry rotation.

Methods:

The UHR CT system features 250×250 µm detector pixels and an x-ray source with a 0.4×0.5 mm focal spot. Visualization of details down to ~150 µm has been reported for this device. A cadaveric femur was imaged on UHR CT at three radial locations within the field-of-view 0 cm (isocenter), 9 cm from the isocenter, and 18 cm from the isocenter; we expect the non-stationary blurs to worsen with increasing radial displacement. Gray level cooccurrence (GLCM) and gray level run length (GLRLM) texture features were extracted from 237 trabecular regions of interest (ROIs, 5 cm diameter) placed at corresponding locations in the femoral head in scans obtained at the different shifts. We evaluated concordance correlation coefficient (CCC) between texture features at 0 cm (reference) and at 9 cm and 18 cm. We also investigated whether the spatially-variant blurs affect K-means clustering of trabecular bone ROIs based on their texture features.

Results:

The average CCCs (against the 0 cm reference) for GLCM and GLRM features were ~0.7 at 9 cm. At 18 cm, the average CCCs were reduced to ~0.17 for GLCM and ~0.26 for GLRM. The non-stationary blurs are incorporated in radiomic features of cancellous bone, leading to inconsistencies in clustering of trabecular ROIs between different radial locations an intersection-over-union overlap of corresponding (most similar) clusters between 0 cm and 9 cm shift was >70%, but dropped to <60% for the majority of corresponding clusters between 0 cm and 18 cm shift.

Conclusion:

Non-stationary CT system blurs reduce inter-scan reproducibility of texture features of trabecular bone in UHR CT, especially for locations >15 cm from the isocenter. Radiomic models of bone quality derived from UHR CT measurements at isocenter might need to be revised before application in peripheral body sites such as the hips.