Twelve-month follow-up results from the SIRONA 2 clinical trial.
Sharif, Faisal; Rosenkranz, Stephen; Bartunek, Jozef; Kempf, Tibor; Aßmus, Birgit; Mahon, Niall G; Hiivala, Nicholas J; Mullens, Wilfried.
Afiliação
  • Sharif F; Department of Cardiology, Galway University Hospital, Saolta Group, CURAM and University of Galway, Galway, Ireland.
  • Rosenkranz S; Clinic III for Internal Medicine, University of Cologne Heart Center and Cologne Cardiovascular Research Center (CCRC), Cologne, Germany.
  • Bartunek J; Cardiovascular Center Aalst, OLV Ziekenhuis Aalst, Aalst, Belgium.
  • Kempf T; Department of Cardiology and Angiology, Hannover Medical School, Hannover, Germany.
  • Aßmus B; Department of Cardiology and Angiology, University Hospital Giessen and Marburg GmbH, Giessen, Germany.
  • Mahon NG; Department of Cardiovascular Medicine, Mater University Hospital and University College Dublin, Dublin, Ireland.
  • Hiivala NJ; Clinical Science, Endotronix Inc, Lisle, IL, USA.
  • Mullens W; Department of Cardiology, Ziekenhuis Oost-Limburg, Genk, Belgium.
ESC Heart Fail ; 11(2): 1133-1143, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38271076
ABSTRACT

AIMS:

In the SIRONA 2 trial, the safety and efficacy of pulmonary artery (PA) pressure (PAP)-guided heart failure (HF) management using a novel PAP sensor were assessed at 30 and 90 days, respectively, and both endpoints were met. The current study examines the prespecified secondary endpoints of safety and accuracy of the PA sensor along with HF hospitalizations and mortality, HF symptoms, functional capacity, quality of life, and patient compliance through 12 months. METHODS AND

RESULTS:

SIRONA 2 is a prospective, multi-centre, open-label, single-arm trial evaluating the Cordella™ PA Sensor System in 70 patients with New York Heart Association (NYHA) functional class III HF with a prior HF hospitalization and/or increase of N-terminal pro-brain natriuretic peptide within 12 months of enrolment. Sensor accuracy was assessed and compared with measurements obtained by standard right heart catheterization (RHC). Safety was defined as freedom from prespecified adverse events associated with use of the Cordella PA Sensor System and was assessed in all patients who entered the cath lab for PA sensor implant. HF hospitalizations and mortality, HF symptoms, functional capacity, quality of life, and patient compliance were also assessed. At 12 months, there was good agreement between the Cordella PA Sensor System and RHC, with the average difference for mean PAP being 2.9 ± 7.3 mmHg. The device safety profile was excellent with 98.4% freedom from device/system-related complications. There were no pressure sensor failures. HF hospitalizations and mortality were low with a rate of 0.33 event per patient year. Symptoms as assessed by NYHA (P < 0.0001) and functional capacity as measured by 6 min walk test (P = 0.02) were significantly improved. Patients' adherence to daily transmissions of PAP and vital signs measurements was 95%.

CONCLUSIONS:

Long-term follow-up of the SIRONA 2 trial supports the safety and accuracy of the Cordella PA Sensor System in enabling comprehensive HF management in NYHA class III HF patients.