Real-Time Detection of Circulating Thrombi in an Extracorporeal Circuit Using Doppler Ultrasound: In-Vitro Proof of Concept Study.
Peer, Syed Murfad; Desai, Manan; Bukhari, Syed; Delores, Yunchuan; Jonas, Richard; Sinha, Pranava.
Afiliação
  • Peer SM; Department of Pediatric Cardiac Surgery, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, USA.
  • Desai M; Department of Cardiovascular Surgery, Children's National Hospital, Washington, DC, USA.
  • Bukhari S; Department of Cardiovascular Surgery, Children's National Hospital, Washington, DC, USA.
  • Delores Y; Department of Laboratory Medicine and Hematology, Children's National Hospital, Washington, DC, USA.
  • Jonas R; Department of Cardiovascular Surgery, Children's National Hospital, Washington, DC, USA.
  • Sinha P; Department of Pediatric Cardiac Surgery, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, USA.
World J Pediatr Congenit Heart Surg ; 15(4): 446-452, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38291667
ABSTRACT

BACKGROUND:

Thromboembolic stroke continues to be by far the most common severe adverse event in patients supported with mechanical circulatory assist devices. Feasibility of using Doppler ultrasound to detect circulating thrombi in an extracorporeal circuit was investigated.

METHODS:

A mock extracorporeal circulatory loop of uncoated cardiopulmonary bypass tubing and a roller pump was setup. A Doppler bubble counter was used to monitor the mean ultrasound backscatter signal (MUBS). The study involved two sets of experiments. In Scenario 1, the circuit was sequentially primed with human blood components, and the MUBS was measured. In Scenario 2, the circuit was primed with heparinized fresh porcine blood, and the MUBS was measured. Fresh blood clots (diameter <1,000 microns, 1,000-5,000 microns, >5,000 microns) were injected into the circuit followed by protamine administration.

RESULTS:

In Scenario 1 (n = 3), human platelets produced a baseline MUBS of 1.5 to 3.5 volts/s. Addition of packed human red blood cells increased the baseline backscatter to 17 to 21 volts/s. Addition of fresh frozen plasma did not change the baseline backscatter. In Scenario 2 (n = 5), the blood-primed circuit produced a steady baseline MUBS. Injection of the clots resulted in abrupt and transient increase (range 3-30 volts/s) of the baseline MUBS. Protamine administration resulted in a sustained increase of MUBS followed by circuit thrombosis.

CONCLUSIONS:

Doppler ultrasound may be used for real-time detection of circulating solid microemboli in the extracorporeal circuit. This technology could potentially be used to design safety systems that can reduce the risk of thromboembolic stroke associated with mechanical circulatory support therapy.