Calibration improvements expand filterscope diagnostic use.
Herfindal, J L; Unterberg, E A; Davda, K M; Garren, E W; Groth, M; Scotti, F; Sontag, A C; Truong, D D; Wilcox, R S.
Afiliação
  • Herfindal JL; Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA.
  • Unterberg EA; Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA.
  • Davda KM; University of Tennessee, Knoxville, Tennessee 37996, USA.
  • Garren EW; Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA.
  • Groth M; Aalto University, Espoo, Finland.
  • Scotti F; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Sontag AC; University of Wisconsin Madison, Madison, Wisconsin 53706, USA.
  • Truong DD; Sandia National Laboratory, Livermore, California 94550, USA.
  • Wilcox RS; Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA.
Rev Sci Instrum ; 95(2)2024 Feb 01.
Article em En | MEDLINE | ID: mdl-38350477
ABSTRACT
The filterscope diagnostic on DIII-D utilizes photomultiplier tubes to measure visible light emission from the plasma. The system has undergone a substantial upgrade since previous attempts to cross-calibrate the filterscope with other spectroscopic diagnostics were unsuccessful. The optics now utilize a dichroic mirror to initially split the light at nearly 99% transmission or reflectance for light below or above 550 nm. This allows the system to measure Dα emission without degrading visible light emission from the plasma for wavelengths below 550 nm (to measure Dß, Dγ, W-I, C-III, etc.). Additional optimization of the optical components and calibration techniques reduce the error in the signal up to 10% in some channels compared to previous methods. Cross-calibration measurements with two other high resolution spectroscopic diagnostics now show excellent agreement for the first time. This expands the capabilities of the filterscope system allowing measurement of divertor detachment, emission profiles, edge-localized mode behavior, and plasma-wall interactions. It also enables direct comparisons against calculations from boundary plasma simulations. These were not possible before.