Experimental periprosthetic fractures with collarless polished tapered cemented stems.
Int Orthop
; 48(5): 1171-1178, 2024 May.
Article
em En
| MEDLINE
| ID: mdl-38443715
ABSTRACT
PURPOSE:
After cemented total hip arthroplasty, the risk of periprosthetic fracture (PPF) of taper-slip stems is higher than that of composite-beam stems. We aimed to assess the conditions resulting in PPFs of taper-slip stems using a falling weight.METHODS:
Taper-slip stems were fixed to five types of simulated bone models using bone cement, and the fractures were evaluated by dropping stainless-steel weights from a predetermined height onto the heads. The periprosthetic fracture height in 50% of the bone models (PPFH50) was calculated using the staircase method.RESULTS:
For the fixation with 0° of flexion, the values for PPFH50 were 61 ± 11, 60 ± 13, above 110, 108 ± 49, and 78 ± 12 cm for the cobalt-chromium-molybdenum alloy, stainless steel alloy (SUS), titanium alloy (Ti), smooth surface, and thick cement mantle models, respectively; for the fixation with 10° of flexion (considering flexure), the PPFH50 values were 77 ± 5, 85 ± 9, 90 ± 2, 89 ± 5, and 81 ± 11 cm, respectively. The fracture rates of the polished-surface stems were 78.6 and 35.7% at the proximal and distal sites, respectively (p < 0.05); the fracture rates of the smooth-surface stems were 14.2 and 100%, respectively (p < 0.05).CONCLUSION:
The impact tests demonstrated that the conditions that were less likely to cause PPFs were use of Ti, a smooth surface, a thick cement mantle, and probably, use of SUS.Palavras-chave