Examining the typical hemodynamic performance of nearly 3000 modern surgical aortic bioprostheses.
Eur J Cardiothorac Surg
; 65(5)2024 May 03.
Article
em En
| MEDLINE
| ID: mdl-38710669
ABSTRACT
OBJECTIVES:
The objective of this analysis was to assess the normal haemodynamic performance of contemporary surgical aortic valves at 1 year postimplant in patients undergoing surgical aortic valve replacement for significant valvular dysfunction. By pooling data from 4 multicentre studies, this study will contribute to a better understanding of the effectiveness of surgical aortic valve replacement procedures, aiding clinicians and researchers in making informed decisions regarding valve selection and patient management.METHODS:
Echocardiograms were assessed by a single core laboratory. Effective orifice area, dimensionless velocity index, mean aortic gradient, peak aortic velocity and stroke volume were evaluated.RESULTS:
The cohort included 2958 patients. Baseline age in the studies ranged from 70.1 ± 9.0 to 83.3 ± 6.4 years, and Society of Thoracic Surgeons risk of mortality was 1.9 ± 0.7 to 7.5 ± 3.4%. Twenty patients who had received a valve model implanted in fewer than 10 cases were excluded. Ten valve models (all tissue valves; n = 2938 patients) were analysed. At 1 year, population mean effective orifice area ranged from 1.46 ± 0.34 to 2.12 ± 0.59 cm2, and dimensionless velocity index, from 0.39 ± 0.07 to 0.56 ± 0.15. The mean gradient ranged from 8.6 ± 3.4 to 16.1 ± 6.2 mmHg with peak aortic velocity of 1.96 ± 0.39 to 2.65 ± 0.47 m/s. Stroke volume was 75.3 ± 19.6 to 89.8 ± 24.3 ml.CONCLUSIONS:
This pooled cohort is the largest to date of contemporary surgical aortic valves with echocardiograms analysed by a single core lab. Overall haemodynamic performance at 1 year ranged from good to excellent. These data can serve as a benchmark for other studies and may be useful to evaluate the performance of bioprosthetic surgical valves over time. CLINICAL TRIAL REGISTRATION NUMBER NCT02088554, NCT02701283, NCT01586910 and NCT01531374.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Valva Aórtica
/
Bioprótese
/
Próteses Valvulares Cardíacas
/
Implante de Prótese de Valva Cardíaca
/
Hemodinâmica
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article