Your browser doesn't support javascript.
loading
Comparison of a portable, pneumotach flow-sensor-based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients-a randomized, crossover study.
Talwar, Deepak; Balamurugan, S; Modi, Mahavir; Salvi, Sundeep; Lopez, Meena; Shah, Rushika; Vaidya, Abhijit; Barne, Monica; Madas, Sapna; Kulkarni, Nandan; Sawant, Sandesh; Gogtay, Jaideep.
Afiliação
  • Talwar D; Department of Pulmonology, Pulmonary Sleep & Critical Care, Metro Centre for Respiratory Diseases, WASOG Sarcoid Clinic, Noida, Uttar Pradesh, India.
  • Balamurugan S; Department of Pulmonology, Chest & Diabetes Research Institute, Chennai, Tamil Nadu, India.
  • Modi M; Department of Pulmonology, Ruby Hall Clinic, Pune, Mumbai, Maharashtra, India.
  • Salvi S; Pulmocare Research & Education Foundation (PURE), Pune, Maharashtra, India.
  • Lopez M; Department of Medical Affairs, Cipla Ltd, Mumbai, Maharashtra, India. meena@cipla.com.
  • Shah R; Department of Medical Affairs, Cipla Ltd, Mumbai, Maharashtra, India.
  • Vaidya A; Department of Medical Affairs, Cipla Ltd, Mumbai, Maharashtra, India.
  • Barne M; Training Research Programs, Chest Research & Training Pvt Ltd, Pune, Maharashtra, India.
  • Madas S; Department of Data Management & Statistics, Chest Research & Training Pvt Ltd, Pune, Maharashtra, India.
  • Kulkarni N; CiplaMedTech, Cipla Ltd, Mumbai, Maharashtra, India.
  • Sawant S; Department of Medical Affairs, Cipla Ltd, Mumbai, Maharashtra, India.
  • Gogtay J; Department of Medical Affairs, Cipla Ltd, Mumbai, Maharashtra, India.
BMC Pulm Med ; 24(1): 230, 2024 May 10.
Article em En | MEDLINE | ID: mdl-38730359
ABSTRACT

BACKGROUND:

Spirofy™ is India's first portable, pneumotach flow-sensor-based digital spirometer developed to diagnose asthma and chronic obstructive pulmonary disease (COPD). In this study, we compared the performance of the Spirofy™ device with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and COPD patients. We also assessed the inter-device variability between two Spirofy™ devices.

METHODS:

In a randomized, three-way crossover, open-label study, we measured the differences in forced expiratory volume in the first second (FEV1) and forced vital capacity (FVC) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. A proportion of the FEV1/FVC ratio distribution of < 0.7 was used to compare the diagnostic accuracies of the Spirofy™ with Vitalograph™ Alpha Touch™ spirometers.

RESULTS:

Ninety subjects participated in this study. The mean ± SD FVC values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touchdevices were 2.60 ± 1.05 L, 2.64 ± 1.04 L, and 2.67 ± 1.04 L, respectively. The mean ± SD FEV1 values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touchdevices were 1.87 ± 0.92 (L), 1.88 ± 0.92 (L), and 1.93 ± 0.93 (L), respectively. A significant positive correlation was found between the FVC and FEV1 values recorded by Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2. As compared to Vitalograph Alpha Touch™, the Spirofy™ device showed good sensitivity (97%), specificity (90%), and overall accuracy (93.3%) at an FEV1/FVC ratio < 0.7. No inter-device variability was observed between the two Spirofy™ devices.

CONCLUSION:

Spirofy™ is a portable and easy-to-use device and is as accurate as the standard Vitalograph Alpha Touch™ spirometer for the diagnosis of COPD and asthma. TRIAL REGISTRATION CTRI/2021/09/036492 (Clinical Trials Registry - India).
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Asma / Espirometria / Estudos Cross-Over / Doença Pulmonar Obstrutiva Crônica Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Asma / Espirometria / Estudos Cross-Over / Doença Pulmonar Obstrutiva Crônica Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Ano de publicação: 2024 Tipo de documento: Article