Leadless pacemaker implementation at the right atrial appendage apex: An initial preclinical assessment.
Lin, Yu-Sheng; Wu, Lung-Sheng; Ho, Wan-Chun; Lai, Chao-Sung; Su, Wilber; Chu, Pao-Hsien.
Afiliação
  • Lin YS; Department of Cardiology, Chang Gung Memorial Hospital, Chiayi, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan., Chang Gung Memorial Hospital, Chiayi, Taiwan.
  • Wu LS; Division of Cardiology, Department of Internal Medicine, Chang Gung Memorial Hospital, Linkou, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan., Chang Gung Memorial Hospital, Taoyuan, Taiwan.
  • Ho WC; Department of Cardiology, Chang Gung Memorial Hospital, Chiayi, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan., Chang Gung Memorial Hospital, Chiayi, Taiwan.
  • Lai CS; Department of Electronics Engineering, Chang Gung University, Taoyuan, Taiwan.
  • Su W; Internal Medicine, Banner-University Medical Center Phoenix, Phoenix, USA.
  • Chu PH; Division of Cardiology, Department of Internal Medicine, Chang Gung Memorial Hospital, Linkou, Taoyuan, Taiwan; College of Medicine, Chang Gung University, Taoyuan, Taiwan., Chang Gung Memorial Hospital, Taoyuan, Taiwan.
Pacing Clin Electrophysiol ; 47(8): 1124-1127, 2024 08.
Article em En | MEDLINE | ID: mdl-38809030
ABSTRACT

OBJECTIVE:

This study evaluates the feasibility and efficacy of implanting a leadless pacemaker at the right atrial appendage (RAA) in a preclinical minipig model, aiming to address the limitations of atrial pacing with current leadless devices like the Medtronic Micra, which is typically used for right ventricular implantation.

METHODS:

Four minipigs, each with a median body weight of 45.8 ± 10.0 kg, underwent placement of the Micra transcatheter pacing system (TPS) via the right femoral vein into the RAA apex. The pacing performance was assessed over 1-week (short-term) and 3-month (long-term) periods.

OUTCOMES:

The initial findings indicated successful implantation, with satisfactory intrinsic R-wave amplitudes and pacing threshold. In the following period, the sensitivity, threshold, and impedance were stable with time. Notably, upon explanation at 3 months, a deep myocardial penetration by the device was observed, necessitating a redesign for safe long-term use in a growing subject's heart.

CONCLUSION:

While initial results suggest that RAA apex placement of the Micra TPS is promising for potential inclusion in a dual-chamber pacing system, the issue of myocardial penetration highlights the need for device redesign to ensure safety and effectiveness in long-term applications.