Accuracy of 12 Intraocular Lens Power Formulas After Corneal Myopic Refractive Surgery.
J Refract Surg ; 40(6): e354-e361, 2024 May.
Article em En | MEDLINE | ID: mdl-38848053
ABSTRACT

PURPOSE:

To assess the predictive accuracy of new-generation online intraocular lens (IOL) power formulas in eyes with previous myopic laser refractive surgery (LRS) and to evaluate the influence of corneal asphericity on the predictive accuracy.

METHODS:

The authors retrospectively evaluated 52 patients (78 eyes) with a history of laser in situ keratomileusis (LASIK) or photorefractive keratectomy (PRK) who subsequently underwent cataract surgery. Refractive prediction errors were calculated for 12 no-history new online formulas 8 formulas with post-LRS versions (Barrett True-K, EVO 2.0, Hoffer QST, and Pearl DGS) using keratometry and posterior/total keratometry measured by IOLMaster 700 and 4 formulas without post-LRS versions (Cooke K6 and Kane) using keratometry and total keratometry. The refractive prediction error, mean absolute error (MAE), and percentages of eyes with prediction errors of ±0.25, ±0.50, ±0.75, ±1.00, and ±1.50 diopters (D) were compared.

RESULTS:

The MAEs of the 12 formulas were significantly different (F = 83.66, P < .001). The MAEs ranged from 0.62 to 0.94 D and from 1.07 to 1.84 D in the formulas with and without post-LRS versions, respectively. The EVO formula produced the lowest MAE (0.60) and MedAE (0.47), followed by the Barrett True-K (0.69 and 0.50, respectively). Each percentage of eyes with refractive prediction error was also significantly different among the 12 formulas (P < .001).

CONCLUSIONS:

The EVO and Barrett True-K formulas demonstrate comparable performance to the other existing formulas in eyes with a history of myopic LASIK/PRK. Surgeons should use these formulas with post-LRS versions and input keratometric values whenever possible. [J Refract Surg. 2024;40(6)e354-e361.].
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Acuidade Visual / Ceratectomia Fotorrefrativa / Implante de Lente Intraocular / Ceratomileuse Assistida por Excimer Laser In Situ / Óptica e Fotônica / Lentes Intraoculares / Miopia Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Acuidade Visual / Ceratectomia Fotorrefrativa / Implante de Lente Intraocular / Ceratomileuse Assistida por Excimer Laser In Situ / Óptica e Fotônica / Lentes Intraoculares / Miopia Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article