Comprehensive study of retention influencing gas chromatographic parameters affecting linear retention indices.
J Chromatogr A
; 1729: 465052, 2024 Aug 16.
Article
em En
| MEDLINE
| ID: mdl-38852268
ABSTRACT
Retention in gas chromatographic systems has a central role in the identification of compounds even if detectors providing spectral information are used. But linear retention indices (LRI) of a single compound originating from multiple sources tend to vary greatly, probably due to differences in the experimental settings of the determinations. The effect of gas chromatographic parameters on LRI has been investigated using 41 compounds - previously identified from food contact plastics - and n-alkanes (n-C7-n-C40) used as reference series. As the reproducibility of LRIs under the same conditions is generally very good, the smallest changes in the settings often caused statistically significant, though irrelevant changes in the LRI values. Therefore, a multicriterial scoring-ranking system has been worked out to highlight the LRI value differences. Our results highlight that column length, heating rate, and film thickness can all be the reasons of the varying published LRI values. We also demonstrated that for the reproduction of LRI data, the chemistry (and not simply the polarity) of the stationary phase is crucial.